Condition Errors 67 to 95, Calibration Failure (Extended)

Errors with instrument internal alignment routines failing.
An event with the error number shown in the table means the condition has been
detected.
When the condition is cleared, an event with the error number plus 1000 is
generated.
For example, error 67 indicates that the RF Alignment not being done during
alignments has been detected, error 1067 indicates that failure has been cleared.
An E in the Error or Warning column means that an error is put up on the front
panel and sent out to SCPI when this condition is detected. A W in this column
means that a Warning is put up on the front panel but nothing goes out to SCPI.
Nothing in this column means nothing is put up; status bit only.
Err#
Message
67
Align 9kHz to 30MHz failed
69
Align 30MHz to 1GHz failed
71
Characterize Preselector
failure
73
unused bit4 is set
75
unused bit4 is set
77
unused bit5 is set
79
unused bit6 is set
81
unused bit7 is set
83
unused bit8 is set
85
unused bit9 is set
87
unused bit10 is set
89
unused bit11 is set
91
unused bit12 is set
93
unused bit13 is set
95
unused bit14 is set
Chapter 3
Condition Errors 1 to 99, Calibration
Error or
Verbose/Correction Information
Warning
W
W
W
The preselector characterization routine
failed.