Calibration
Full 4-Port Calibration

Procedure

Figure 4-17
Connecting the Standard at Full 4-Port Calibration
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Connect an OPEN calibration standard to test port 1 (the connector to which the DUT will
be connected).
Step 6. Press
Step 7. Disconnect the OPEN calibration standard connected in Step 5 and replace it with a
SHORT calibration standard.
Step 8. Press
Step 9. Disconnect the SHORT calibration standard connected in Step 7 and replace it with a
LOAD calibration standard.
Step 10. Press
Step 11. Repeat Step 5 to Step 10 on test port 2.
Step 12. Repeat Step 5 to Step 10 on test port 3.
Step 13. Repeat Step 5 to Step 10 on test port 4.
100
.
.
Calibrate
.
4-Port Cal
.
Reflection
to start the calibration measurement.
Port 1 Open
to start the calibration measurement.
Port 1 Short
to start the calibration measurement.
Port 1 Load
Chapter 4