Sweep Order in Each Channel

In a channel, each test port is set to stimulus port in the order of port number and updates
each trace (Table 5-1).
Table 5-1
Sweep Order in Each Channel
Sweep Order
*1.Option 313, 314, 413, or 414 only
*2.Option 413 or 414 only
NOTE
If full 2, 3, or 4 port error correction is valid, any traces related to the calibrated ports are
not updated before the last calibrated port is swept as a stimulus port.
Sweep is not executed for the stimulus port which is not required for updating traces.

Trigger Source

The trigger source generates a cue signal that initiates a measurement process. Four types
of trigger sources are available as shown in Table 5-2.
Table 5-2
Trigger Sources and Their Functions
Trigger Sources
Internal (
External (
Manual (
Bus (
Chapter 5
Setting Up the Trigger and Making Measurements
Stimulus Port
Port 1
Port 2
*1
Port 3
*2
Port 4
Function
)
Uses a consecutive signal generated by the firmware as a
Internal
trigger source. Triggers are sent immediately following the
completion of each measurement.
)
Uses the external trigger input terminal (BNC) as a trigger
External
source.
)
Manual
A trigger will be generated by pressing
)
A trigger will be generated by executing a *TRG command.
Bus
Making Measurements
Updated Trace
*1
*2
S
, S
, S
, S
11
21
31
41
*1
*2
S
, S
, S
, S
12
22
32
42
*1
*1
*1
*2
S
, S
, S
, S
13
23
33
43
*2
*2
*2
*2
S
, S
, S
, S
14
24
34
44
-
.