Defining Parameters For Standards - Agilent Technologies ENA Series User Manual

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Defining Parameters for Standards

Figure 4-27 and Figure 4-28 show the parameters used in defining standards.
Figure 4-27
Reflection Standard Model (SHORT, OPEN, or LOAD)
Figure 4-28
Transmission Standard Model (THRU)
Z0
Delay
Chapter 4
Changing the Calibration Kit Definition
The offset impedance between the standard to be defined and the
actual measurement plane. Normally, this is set to the characteristic
impedance of the system.
The delay occurs depending on the length of the transmission line
between the standard to be defined and the actual measurement plane.
In an OPEN, SHOT, or LOAD standard, the delay is defined as
one-way propagation time (sec.) from the measurement plane to the
standard. In a THRU standard, it is defined as one-way propagation
time (sec.) from one measurement plane to the other. The delay can be
Calibration
111

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5070E5071a

Table of Contents