Analysis in Time Domain (Option 010)

Deleting unnecessary data in time domain (gating)

Deleting unnecessary data in time domain (gating)

Flow of measurement

Table 8-1 shows the flow of measurement. Figure 8-8 shows the change of the waveform in
each flow.
Table 8-5
Measurement flow
Item
1. Measurement in frequency
domain
2. Transformation to time domain
3. Setting the gate
4. Transforming the data back to
data in frequency domain
Figure 8-8
Measurement flow

Setting gate type

The E5070A/E5071A lets you choose from the following 2 gate types:
Gate type
Band pass
Notch
188
Description
Execute measurement in frequency domain.
Enable the transformation function and transform measurement data to
data in time domain.
To select a necessary domain, make the following settings of the gate.
Gate type
Gate shape
Gate range
Disable the transformation function and display the response in
frequency domain corresponding to the data selected with the gate.
Description
Deletes the response outside the gate range.
Deletes the response inside the gate range.