Corrected System Performance - Agilent Technologies ENA Series User Manual

Rf network analyzer
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Corrected System Performance

The specifications in this section apply for measurements made with the Agilent
E5070A/E5071A Network Analyzer under the following conditions:
No averaging applied to data
Environmental temperature of 23°C ±5°C, with less than 1°C deviation from the
calibration temperature
Response and isolation calibration not omitted
Table 15-1
System Dynamic Range
Description
*1*2
System Dynamic Range
300 kHz to 3 MHz
IF bandwidth = 3 kHz
3 MHz to 1.5 GHz
1.5 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 7.5 GHz
7.5 GHz to 8.5 GHz
300 kHz to 3 MHz
IF bandwidth = 10 Hz
3 MHz to 1.5 GHz
1.5 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 7.5 GHz
7.5 GHz to 8.5 GHz
*1.The test port dynamic range is calculated as the difference between the test port rms noise floor and the source
maximum output power. The effective dynamic range must take measurement uncertainty and interfering signals
into account.
*2.May be limited to 90 dB at particular frequencies below 350MHz or above 4.2GHz due to spurious receiver resid-
uals.
Chapter 15
Specifications and Supplemental Information
Corrected System Performance
Specification
95 dB
97 dB
95 dB
92 dB
85 dB
120 dB
122 dB
120 dB
117 dB
110 dB
Supplemental Information
85 dB
98 dB
100 dB
97 dB
95 dB
88 dB
110 dB
123 dB
125 dB
122 dB
120 dB
113 dB