Setting Measurement Conditions
Stimulus Settings

Setting the Power Level

The power level can be set to any value from -15 dBm to 0 dBm (-50 dBm to 0 dBm for
models with Option 214, 314, or 414) with a minimum increment of 0.05 dB.
Setting the Power Level
Step 1. Press
set.
Step 2. Press
Step 3. If Option 214, 314, or 414 is installed, follow the procedure below:
a. Press
Power Ranges
b. Press the softkey that corresponds to the desired power range.
Softkey
-15 to 0
-20 to -5
-25 to -10
-30 to -15
-35 to -20
-40 to -25
-45 to -30
-50 to -35
Step 4. Press
.
Power
Step 5. Using the ENTRY block keys on the front panel, input the power level.

Setting the Number of Points

The number of points is the number of data items collected in one sweep. It can be set to
any number from 2 to 1601 for each channel independently.
To obtain a higher trace resolution against the stimulus value, choose a larger number
of points.
To obtain higher throughput, keep the number of points to a smaller value within an
allowable trace resolution.
To obtain higher measurement accuracy after calibration, perform calibration using the
same number of points as in actual measurements.
62
or
to select the channel for which the power level will be
.
.
Function
Sets the power range to -15 dBm to 0 dBm.
Sets the power range to -20 dBm to -5 dBm.
Sets the power range to -25 dBm to -10 dBm.
Sets the power range to -30 dBm to -15 dBm.
Sets the power range to -35 dBm to -20 dBm.
Sets the power range to -40 dBm to -25 dBm.
Sets the power range to -45 dBm to -30 dBm.
Sets the power range to -50 dBm to -35 dBm.
Chapter 3