Step 6. Press the
appear.
Figure 14-10
EcalAssistant (Connection) Dialog Box
Step 7. Connect the ECal module to the test ports according to the instruction in the dialog box and
then press
Setup Description
Executing calibration between ports 1 and 2
Executing calibration between ports 1 and 3
Executing calibration between ports 2 and 3
Figure 14-11
Connecting the ECal Module
Step 8. The EcalAssistant (Complete) dialog box (Figure 14-12) will appear. Press the
button to terminate the calibration.
Chapter 14
button. The EcalAssistant (Connection) dialog box (Figure 14-10) will
Next
.
Measure
Measurement Examples
Evaluating a Duplexer
Key Operation
(After making the connections shown in Figure
14-3 (a)), press
-
.
Measure
Next
(After making the connections shown in Figure
14-3 (b)), press
-
.
Measure
Next
(After making the connections shown in Figure
14-3 (c)), press
-
.
Measure
Next
Done
291