1-Port Calibration (Reflection Test)
In 1-port calibration, calibration data are measured by connecting an OPEN standard, a
SHORT standard, and a LOAD standard to the desired test port. This calibration effectively
eliminates the frequency response reflection tracking error, directivity error, and source
match error from the test setup in a reflection test using that port (Figure 4-10).
Figure 4-10
1-Port Error Model (1-Port Calibration)
Procedure
Figure 4-11
Connecting the Standard at 1-Port Calibration
Chapter 4
1-Port Calibration (Reflection Test)
Calibration
91