Softkey Functions
Calibration Menu
Key Operation
(Continued)
Calibrate
1-Port Cal
Select Port
Returns to the softkey display screen one level higher.
Cancel
Executes an OPEN calibration on the test port selected in the
Open
Select Port
Executes a SHORT calibration on the test port selected in the
Short
Select Port
Executes a LOAD calibration on the test port selected in the
Load
menu.
Port
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels the calibration in progress.
OK
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing a full 2-port calibration.
2-Port Cal
Displays softkeys for selecting test ports.
*1
Select Ports
Selects test ports 1 and 2.
1-2
Selects test ports 1 and 3.
1-3
Selects test ports 1 and 4.
*2
1-4
Selects test ports 2 and 3.
2-3
Selects test ports 2 and 4.
*2
2-4
Selects test ports 3 and 4.
*2
3-4
Returns to the softkey display screen one level higher.
Cancel
Displays softkeys for executing reflection calibrations. Symbols x and y refer to the two test ports selected in the
Reflection
Select Ports
Executes an OPEN reflection calibration on test port x.
Port x Open
Executes a SHORT reflection calibration on test port x.
Port x Short
Executes a LOAD reflection calibration on test port x.
Port x Load
Executes an OPEN reflection calibration on test port y.
Port y Open
Executes a SHORT reflection calibration on test port y.
Port y Short
Executes a LOAD reflection calibration on test port y.
Port y Load
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing transmission calibrations. Symbols x and y refer to the two test ports selected in
Transmission
the
Select Ports
Executes a THRU calibration on test ports x and y.
Port x-y Thru
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing isolation calibrations. Symbols x and y refer to the two test ports selected in the
Isolation
(Optional)
Select Ports
Executes an isolation calibration on test ports x and y.
Port x-y Isol
Returns to the softkey display screen one level higher.
Return
416
Function
menu.
menu.
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
SCPI Command
:SENS{1-9}:CORR:COLL:OPEN
:SENS{1-9}:CORR:COLL:SHOR
:SENS{1-9}:CORR:COLL:LOAD
Select
:SENS{1-9}:CORR:COLL:SAVE
None
:SENS{1-9}:CORR:COLL:METH:SOLT2 1,2
:SENS{1-9}:CORR:COLL:METH:SOLT2 1,3
:SENS{1-9}:CORR:COLL:METH:SOLT2 1,4
:SENS{1-9}:CORR:COLL:METH:SOLT2 2,3
:SENS{1-9}:CORR:COLL:METH:SOLT2 2,4
:SENS{1-9}:CORR:COLL:METH:SOLT2 3,4
:SENS{1-9}:CORR:COLL:OPEN x
:SENS{1-9}:CORR:COLL:SHOR x
:SENS{1-9}:CORR:COLL:LOAD x
:SENS{1-9}:CORR:COLL:OPEN y
:SENS{1-9}:CORR:COLL:SHOR y
:SENS{1-9}:CORR:COLL:LOAD y
:SENS{1-9}:CORR:COLL:THRU x,y
:SENS{1-9}:CORR:COLL:THRU y,x
:SENS{1-9}:CORR:COLL:ISOL x,y
:SENS{1-9}:CORR:COLL:ISOL y,x