Key Operation
Function
(Continued)
Data Math
Multiplies the measured data by the memory trace data and stores the result
Data * Mem
as a data trace.
Subtracts the memory trace data from the measured data (vector operation)
Data - Mem
and stores the results as a data trace. This function is useful in determining
vector errors (e.g., directivity) and storing them for data compensation for
future measurements.
Adds the measured data to the memory trace data and stores the result as a
Data + Mem
data trace.
Returns to the softkey display screen one level higher.
Cancel
Displays the input bar for editing title labels. Each channel is allowed to have
Edit Title Label
a title label up to 100 characters in length. However, the entire title bar may
not be displayed depending on the width of the channel window.
Enables or disables the title label display. Once a title label is assigned, it can
Title Label
be displayed within the title bar at the top of a channel window by enabling
the title label display. Title labels are useful for adding extra information to
saved data and printouts.
Enables or disables the graticule label display for the Y-axis in a rectangular
Graticule Label
display format. Disabling the graticule label display will result in expanded
graph areas.
Inverts the display colors for all channel windows. Inverting colors from the
Invert Color
normal display will create white backgrounds.
Enables or disables the frequency display on the screen. This function is
Frequency
useful for protecting critical information by making it impossible to obtain
frequencies used in measurements just by glancing at the screen.
Enables or disables screen updates. When disabled, the analyzer can save
Update
time required for screen updates and thus improve measurement throughput.
If it is not necessary to confirm results on the screen, this is an effective way
to improve throughput.
Returns to the "E5070A/E5071A Menu (Top Menu)" on page 404.
Return
Appendix D
Softkey Functions
Display Menu
SCPI Command
:CALC{1-9}:MATH:FUNC MULT
:CALC{1-9}:MATH:FUNC SUBT
:CALC{1-9}:MATH:FUNC ADD
:DISP:WIND{1-9}:TITL:DATA
:DISP:WIND{1-9}:TITL
:DISP:WIND{1-9}:LAB
:DISP:IMAG
:DISP:ANN:FREQ:STAT
:DISP:ENAB
427