Optimizing Measurements

Reduce Measurement Error in High Temperature Environments

Reduce Measurement Error in High Temperature
Environments
The E5070A/E5071A is designed to obtain the best measurement accuracy at the ambient
temperature range of 23°C ± 5°C. The high temperature measurement mode of the
E5070A/E5071A reduces measurement error (drift error) at an ambient temperature of
28°C to 33°C.
NOTE
The high temperature measurement mode must be turned off when the analyzer is used at
an ambient temperature below 28°C. Otherwise, the measurement accuracy may be
degraded.

Procedure

Step 1. Press
Step 2. Press
Step 3. Press
234
.
.
Service Menu
to turn
High Temperature
the high temperature measurement mode.
ON/OFF
Chapter 11