Softkey Functions
Calibration Menu
Key Operation
(Continued)
Calibrate
*1
3-Port Cal
Terminates a calibration process and calculates the calibration
Done
coefficients.
Displays softkeys for canceling a calibration.
Cancel
Cancels the calibration in progress.
OK
Returns to the softkey display screen one level higher.
Cancel
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing full 4-port calibrations.
*2
4-Port Cal
Displays softkeys for executing reflection calibrations.
Reflection
Executes an OPEN calibration on test port 1.
Port 1 Open
Executes a SHORT calibration on test port 1.
Port 1 Short
Executes a LOAD calibration on test port 1.
Port 1 Load
Executes an OPEN calibration on test port 2.
Port 2 Open
Executes a SHORT calibration on test port 2.
Port 2 Short
Executes a LOAD calibration on test port 2.
Port 2 Load
Executes an OPEN calibration on test port 3.
Port 3 Open
Executes a SHORT calibration on test port 3.
Port 3 Short
Executes a LOAD calibration on test port 3.
Port 3 Load
Executes an OPEN calibration on test port 4.
Port 4 Open
Executes a SHORT calibration on test port 4.
Port 4 Short
Executes a LOAD calibration on test port 4.
Port 4 Load
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing transmission calibrations.
Transmission
Executes a THRU calibration between test ports 1 and 2.
Port 1-2 Thru
Executes a THRU calibration between test ports 1 and 3.
Port 1-3 Thru
Executes a THRU calibration between test ports 1 and 4.
Port 1-4 Thru
Executes a THRU calibration between test ports 2 and 3.
Port 2-3 Thru
Executes a THRU calibration between test ports 2 and 4.
Port 2-4 Thru
Executes a THRU calibration between test ports 3 and 4.
Port 3-4 Thru
Returns to the softkey display screen one level higher.
Return
Displays softkeys for executing isolation calibrations.
Isolation
(Optional)
Executes an isolation calibration on test ports 1 and 2.
Port 1-2 Isol
Executes an isolation calibration on test ports 1 and 3.
Port 1-3 Isol
Executes an isolation calibration on test ports 1 and 4.
Port 1-4 Isol
Executes an isolation calibration on test ports 2 and 3.
Port 2-3 Isol
418
Function
SCPI Command
:SENS{1-9}:CORR:COLL:SAVE
None
:SENS{1-9}:CORR:COLL:OPEN 1
:SENS{1-9}:CORR:COLL:SHOR 1
:SENS{1-9}:CORR:COLL:LOAD 1
:SENS{1-9}:CORR:COLL:OPEN 2
:SENS{1-9}:CORR:COLL:SHOR 2
:SENS{1-9}:CORR:COLL:LOAD 2
:SENS{1-9}:CORR:COLL:OPEN 3
:SENS{1-9}:CORR:COLL:SHOR 3
:SENS{1-9}:CORR:COLL:LOAD 3
:SENS{1-9}:CORR:COLL:OPEN 4
:SENS{1-9}:CORR:COLL:SHOR 4
:SENS{1-9}:CORR:COLL:LOAD 4
:SENS{1-9}:CORR:COLL:THRU 1,2
:SENS{1-9}:CORR:COLL:THRU 2,1
:SENS{1-9}:CORR:COLL:THRU 1,3
:SENS{1-9}:CORR:COLL:THRU 3,1
:SENS{1-9}:CORR:COLL:THRU 1,4
:SENS{1-9}:CORR:COLL:THRU 4,1
:SENS{1-9}:CORR:COLL:THRU 2,3
:SENS{1-9}:CORR:COLL:THRU 3,2
:SENS{1-9}:CORR:COLL:THRU 2,4
:SENS{1-9}:CORR:COLL:THRU 4,2
:SENS{1-9}:CORR:COLL:THRU 3,4
:SENS{1-9}:CORR:COLL:THRU 4,3
:SENS{1-9}:CORR:COLL:ISOL 1,2
:SENS{1-9}:CORR:COLL:ISOL 2,1
:SENS{1-9}:CORR:COLL:ISOL 1,3
:SENS{1-9}:CORR:COLL:ISOL 3,1
:SENS{1-9}:CORR:COLL:ISOL 1,4
:SENS{1-9}:CORR:COLL:ISOL 4,1
:SENS{1-9}:CORR:COLL:ISOL 2,3
:SENS{1-9}:CORR:COLL:ISOL 3,2
Appendix D