Mounting Requirements; Noise; Radiation Shielding - GE SENOGRAPHE DMR+ Preinstallation Manual

Hide thumbs Also See for SENOGRAPHE DMR+:
Table of Contents

Advertisement

GE Healthcare
REV 1
SECTION 2

MOUNTING REQUIREMENTS

TABLE 3-2
MOUNTING REQUIREMENTS
PRODUCT/
COMPONENT
Gantry
Rad shield screen
SECTION 3

NOISE

SECTION 4

RADIATION SHIELDING

Given that the Senographe DMR has a modular design, its optimal use depends on the relative
position of the modules (see the Installation Plan prepared for each site).
However, the gantry and rad shield screen require a minimum amount of "free space" (see
Table 3–2 below).
Length
mm (feet)
3000
(10)
1525
(5)
D
60 dBA at 1 m (3'3" ).
The RAD shield screen has a 1 mm (0.04") lead thickness equivalency.
The limit value for the whole–body equivalent dose must not exceed 50 mSev (5Rem) per
year for the European Regulations.
The minimum permissible distances to be out of these limits, under the mentioned parameters,
are given on page 3–12/Illustration 3–9.
3–11
SENOGRAPHE DMR+
Width
mm (feet)
2500
(8.20)
1200
(3.94)
pim 5128706–1–100
Height
mm (feet)
2500
(8.20)
2000
(6.56)