Sterilization Of Exp V. Monoblock (Exhalation Block With Flow Sensor) - Siare Aria 104 User Manual

Lung ventilator
Table of Contents

Advertisement

6.3.1 Sterilization of EXP V. Monoblock (exhalation block with flow sensor)

EXP V.
Monoblock
6-8
Disinfect with steam (93° - 10 minutes) or
chemically.
CAUTION
The EXP V. monoblock includes the expiratory valve and the flow
sensor.
Do not attempt to dismantle or clean with compressed air.
The EXP. V. monoblock can be washed and disinfected by immersing it
in a bowl with 3 centimetres of liquid (chemical), keeping the connector
for the electrical connections facing upwards, following the
manufacturer's instructions (see following note about recommended
chemical agents).
The following are chemical agents recommended for cold sterilization
of flow sensor.
5.25% - 6.15% Sodium Hypochlorite solution
Protex wipes (Didecyl Dimethyl Ammonium Chloride)
WARNING !!
Risk of device failure
EXP valve and EXP flow sensor disinfection.
Do not use formaldehyde or phenol-based disinfectants as they can
cause cracking and reticulation of plastic parts.
Do not use too strong disinfectants as they can compromise the
working life of the immersed parts.
Rinsing and carefully drying the components since marks and other
damage can occur when the components are exposed to high
temperatures.
It is possible to sterilize the
component with gamma rays
or ethylene oxide (ETO).
When disinfection is complete,
rinse with running, preferably
decalcified, water; shake and
drain off any remaining water.
Leave the components to dry
completely.
User manual, DU3104101

Advertisement

Table of Contents
loading

Table of Contents