5.12.3 Respiratory Flow Sensors Calibration
The Respiratory Flow Sensors Calibration is necessary in the following cases.
•
When the User note differences are more than 15% (over 100ml) between the set Volume
value ( VTi - Vte ) and the expired Tidal Volume reading ( Vte ).
•
In case of first calibration of ventilation module or replacement of flow sensors (INSP inside the
unit or EXP outside the unit), it is suggested to perform this calibration.
In order to start the Flow Sensors Calibration is necessary the intervention of
qualified SIARE personnel or qualified technical personnel authorized by SIARE. For
more information on the procedure please refer to the SERVICE manual.
Morpheus_ND
•
Select Respiratory Flow Sensors
Calibration.
The page to perform Expiratory Flow
Sensors Calibration is displayed.
•
Select
The Expiratory Flow Sensors
Calibration program is performed.
•
Select
The Expiratory Flow Sensors
Calibration phase may be aborted.
•
Select
The User can check the correct
calibration through the displaying of
the flow parameters (l/min)
Inspiratory and Expiratory Flow.