5.3.8 Turbine calibration
Follow the indications in chapter 6 : Calibration
•
•
•
Press the soft keys on the ventilator keyboard at the same time : ALARM RESET
and ON-OFF keys.
5.3.9 Flow sensor calibration
Follow the indications in chapter 6 : Calibration
•
•
•
Press the soft keys on the ventilator keyboard at the same time : ALARM RESET
and ON-OFF keys.
5.3.10 Trigger
•
VC/VAC
operating mode
Press the AUT key, make sure that the PAW values displayed on the flow
analyser and on the ventilator fall within 10 and 20 cmH 2 O.
Siaretron 4000 OFF
Patient circuit installed and plugged
Flow sensor and oxygen sensor installed
If the device does not exceed the SELF TEST please follow the
instructions in chapter 6 : Calibration
Siaretron 4000 OFF
Patient circuit installed and plugged
Flow sensor and oxygen sensor installed
If the device does not exceed the SELF TEST please follow the
instructions in chapter 6 : Calibration
Patient circuit installed and connected to the lung simulator for adults.
The Current Volume values displayed on the flow analyser (Vti)
and on the ventilator monitored parameters (Vte), must fall within
the relative acceptance limits (e.g. 500 ml +/- 15%).