Summary Table Of Alarm Characteristics; Alarms Configurable By User - Siare Falco 202 Evo User Manual

Lung ventilator
Table of Contents

Advertisement

5.3

Summary table of alarm characteristics

5.3.1 Alarms configurable by user

Alarms
Low Pressure of
Airways
High Pressure of
Airways
Low Respiratory
Rate
High Respiratory
Rate
Low Expiratory
Volume
High Expiratory
Volume
Low Volume
Minute
High Volume
Minute
Low PEEP
High PEEP
Low O₂
Concentration
High O₂
Concentration
Low EtCO
2
High EtCO
2
Apnoea
On Battery
5-16
Delay Time
Priority
(s)
HIGH
15
HIGH
0
3 patient
HIGH
breaths
3 patient
HIGH
breaths
HIGH
15
HIGH
0
HIGH
60
HIGH
60
HIGH
0
HIGH
0
HIGH
30
HIGH
30
HIGH
30
HIGH
30
HIGH
5 - 60
HIGH
15
Suspendable
Suspendable
for (s)
YES
NO
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
Inhibition
15
NO
-
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
30
NO
120
YES
User manual, DU3104101