Test Table - Siare Siaretron 4000 Service Manual

Icu ventilator, turbine-driven ventilation
Hide thumbs Also See for Siaretron 4000:
Table of Contents

Advertisement

5.4

Test table

Ref.
Check
Aesthetics
1
Presence of markings
2
3
Software version
SELF TEST
4
Turbine
Acoustic alarm
5
Patient circuit loss (TEST)
O 2 sensor calibration (TEST)
6
O 2 sensor check
7
8
Turbine calibration
Flow Sensor Calibration
9
Flow TRIGGER check
10
Pressure TRIGGER check
Respiratory
rates check
11
Low/high respiratory rate alarm
Low pressure
12
circuit loss
13
Power supply alarm
Gas supply alarm
14
INSP HOLD Function
15
16
EXP HOLD Function
Nebulizer ( 6l/min )
17
Function 100% O 2
18
Measured values
Rel. ________
_____ mV
(21%)
_____ %
_____ %
(100%)
Sett. 12 bpm
Meas. _____
Sett. 40 bpm
Meas. _____
Sett. 70 bpm
Meas. _____
Sett. 21%
Sett. 100 %
Meas. _____
Result
Notes
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.
 Pos.
 Neg.