Output display

The system meets the AIUM (American Institute of Ultrasound in Medicine) output display standard for MI
and TI (see
"Related guidance documents"
transducer and operating mode when either the TI or MI is greater than or equal to a value of 1.0, thus
requiring display.
Table 11: TI or MI ≥ 1.0
Transducer
Index
Model
C11xp
MI
TIC, TIB, or TIS
C35xp
MI
TIC, TIB, or TIS
C60xp
MI
TIC, TIB, or TIS
D2xp
MI
TIC, TIB, or TIS
HFL38xp
MI
TIC, TIB, or TIS
HFL50xp
MI
TIC, TIB, or TIS
HSL25xp
MI
TIC, TIB, or TIS
Even if MI is less than 1.0, the system provides a continuous real-time display of MI in all imaging modes, in
increments of 0.1.
The system meets the output display standard for TI and provides a continuous real-time display of TI in all imaging
modes, in increments of 0.1.
The TI consists of three user-selectable indices, and only one of these is displayed at any one time. In order to
display TI properly and meet the ALARA principle, the user selects an appropriate TI based on the specific exam
being performed. FUJIFILM SonoSite provides a copy of
guidance on determining which TI is appropriate (see
Acoustic output
on page 10-9). The following table indicates for each
2D/ M Mode
CPD/Color
No
No
No
No
Yes
No
Yes
No
Yes
Yes
Yes
Yes
-
-
-
-
Yes
Yes
No
No
Yes
Yes
No
No
Yes
Yes
No
No
AIUM Medical Ultrasound Safety
"Related guidance documents"
PW Doppler
CW Doppler
Yes
-
Yes
-
Yes
-
Yes
-
Yes
-
Yes
-
-
No
-
Yes
No
-
Yes
-
No
-
Yes
-
No
-
Yes
-
, which contains
on page on page 10-9).