9 Allow the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated
space.
10 Examine the system, transducer and cable for damage, such as cracks or splitting where fluid can enter.
If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local
representative.

Storing the transducer

To store the transducer
1 Make sure the transducer has been cleaned and disinfected as detailed in the previous section.
2 Store the transducer so that it hangs freely and vertically, and observe the following precautions:
Store the transducer away from any contaminated transducers.
Store the transducer in an environment that is safe and has good airflow. Do not store the transducer
in closed containers or where condensation may occur.
Avoid direct sunlight and exposure to x-rays. Recommended storage temperature range is between
o
o
0
C (32
F) and +45
If using a wall-mounted rack for storage, ensure that:
It is securely mounted.
The storage slots do not mar the transducer or cause damage to the cable.
The rack is sized and positioned to prevent the transducer from inadvertently falling.
Make sure the connector is supported and secure.

Transporting the transducer

When transporting the transducer, you must take precautions to protect the transducer from damage and
avoid cross-contamination. Be sure to use a container approved by your organization.
To transport a soiled transducer for cleaning
A soiled transducer is one that has been contaminated and must be cleaned before using it in an exam.
Cleaning and disinfecting
o
o
C (113
F).
8-11