8 Verify that all gel, debris, and bodily fluids have been removed from the system and transducer. If
necessary, repeat steps 5, 6, and 7 with a new wipe.
WARNING
9 Prepare the disinfectant for use.
a Choose a high-level disinfectant from the list of approved disinfectants.
High-level disinfectants compatible with SonoSite X-Porte transducers:
a
Compatible transducers
Disinfectant
Cidex
C11xp, C35xp, C60xp, HFL38xp,
HFL50xp, HSL25xp, ICTxp,
L25xp, L38xp, P10xp, P21xp,
rP19xp
Cidex OPA
C11xp, C60xp, HFL38xp,
HFL50xp, HSL25xp, ICTxp,
L25xp, L38xp, P10xp, P21xp,
rP19xp
Revital-Ox
C35xp
RESERT
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.
b Check the expiration date on the bottle to ensure the disinfectant has not expired.
c Mix or check that the disinfection chemicals have the concentration recommended by the manufacturer
(for example, a chemical strip test).
d Check that the temperature of the disinfectant is within the manufacturer's recommended limits.
10 Perform a high-level disinfection of the transducer. Use the following procedure:
a Immerse the transducer in a high-level disinfectant solution.
WARNING
8-6
Failure to remove all gel, debris, and bodily fluid could leave contaminants on the
transducer.
High-level disinfectants can cause harm to the patient if not completely removed
from the transducer. Follow the manufacturer's rinse instructions to remove
chemical residue.
Temperature
Disinfectant soak duration
o
o
45 minutes
25
C, 77
F
o
o
12 minutes
20
C, 68
F
o
o
8 minutes
20
C, 68
F
www.sonosite.com
for a complete list of the most
Cleaning and disinfecting