Gynecological Calculations - FujiFilm SonoSite X-Porte User Manual

Hide thumbs Also See for SonoSite X-Porte:
Table of Contents

Advertisement

To calculate Tricuspid Annular Plane Systolic Excursion (TAPSE)
TAPSE is used to determine the state of the right ventricular systolic function and to correlate this
measurement with the right ventricular ejection fraction.
1 On a frozen M Mode sweep, tap Calcs.
If this control is hidden, tapping More Controls displays it.
2 On the calculation list, tap TAPSE.
3 On the measurements list, tap TAPSE.
4 Position the calipers by dragging.
5 To save the measurements, tap
To measure a Tissue Doppler Imaging (TDI) waveform
1 Ensure that TDI is on (see
2 On a frozen Doppler spectral trace, tap Calcs.
If this control is hidden, tapping More Controls displays it.
3 Do the following for each measurement you want to take:
a On the TDI calculation list, tap the measurement name.
b Position the calipers by dragging.
c Save the calculation: Tap

Gynecological calculations

WARNINGS
The Gynecological calculations include measurements for uterus, ovaries, and follicles.
In the Gynecological exam, you can also calculate volume (see
Measurements and calculations
under the measurement name.
"Controls in Doppler"
under the measurement name.
To avoid incorrect calculations, verify that the patient information, date, and
time settings are accurate.
To avoid misdiagnosis or harming the patient outcome, start a new patient
form before starting a new patient exam and performing calculations. Starting
a new patient form clears the previous patient's data. The previous patient's
data will be combined with the current patient if the form is not first cleared.
See
"Editing patient information"
on page 4-16).
on page 4-31.
"Volume calculation"
on page 5-12).
5-35

Hide quick links: