To calculate Proximal Isovelocity Surface Area (PISA)
The PISA calculation requires a measurement in 2D, a measurement in Color, and two measurements in
Doppler spectral trace. After all measurements are saved, the result appears in the patient report.
1 Measure from Ann D:
a On a frozen 2D image, tap Calcs.
If this control is hidden, tapping More Controls displays it.
b On the PISA calculation list, tap Ann D.
c Position the calipers by dragging.
d Save the calculation: Tap
2 Measure from Radius:
a On a frozen Color image, tap Calcs.
b On the PISA calculation list, tap Radius.
c Position the calipers by dragging.
d Save the calculation: Tap
3 On a frozen Doppler spectral trace, tap Calcs.
4 Do the following to measure from MR VTI and again to measure from MV VTI:
a On the PISA calculation list, tap MR VTI or MV VTI.
b Drag the caliper to the start of the waveform, and lift your finger to start the trace.
A pencil icon appears.
c Using your finger, trace the waveform.
To make a correction, trace backward on the dotted line.
d Lift your finger to complete the trace.
Note
e Save the calculation: Tap
To measure LVDd and LVDs
You may use these measurements to calculate LV volume and ejection fraction (EF).
5-24
under the measurement name.
under the measurement name.
Even after completing the trace, you can still adjust it by dragging the cursor
backward on the dotted line.
under the measurement name.