Cautions
b Ensure that the connector and most of the cable remain out of the fluid. You may submerge 2 inches
(5 cm) of the cable proximal to the transducer scanhead.
Connector
Do not submerge
Figure 8-1
Transducer component names
11 Rinse the transducer 3 separate times using the following procedure:
a Rinse the transducer in clean, running water according to the disinfectant manufacturer's instructions (at
least 1 minute).
b Ensure that the connector and at least 12-18 inches (31-46 cm) of the cable from the connector remain
out of the fluid.
c Replace the rinse water before beginning the next rinse.
12 Dry the transducer with a sterile lint-free cloth.
13 Dispose of the disinfectant according to the manufacturer's guidelines.
14 Examine the transducer and cable for damage, such as cracks or splitting where fluid can enter.
If damage is evident, discontinue use of the transducer, and contact FUJIFILM SonoSite or your local
representative.
Cleaning and disinfecting
Do not soak the transducer longer than recommended by the chemical
manufacturer.
Do not immerse the transducer connector in any disinfectant solution.
Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a
non- recommended disinfecting solution or incorrect solution strength can
damage or discolor the transducer and void the warranty.
Cable
Maximum
5 cm
(2 inches)
Scan head
8-7