To remove a transducer
Caution
1 Pull the transducer latch up, and rotate it clockwise.
2 Pull the transducer connector away from the system.

Selecting a transducer and exam type

WARNING
Before scanning, select a transducer and exam type.
To select a transducer and exam type
1 Do one of the following:
On the start select screen, tap SELECT if present. To set up the start select screen, see
settings"
on page 3-32.
Tap TRANSDUCERS & EXAMS at the top of the touch panel.
Cards for the available transducers appear (see
2 On the card for the appropriate transducer, do one of the following:
Double-tap the exam type.
Tap the exam type, and then tap SCAN, or tap Cancel to cancel.
Scrolling the list of exam types displays any hidden items.
You can also select the exam type on the patient form (see
page 4-31).
Related topics
Intended uses ........................................................................................................................................................... 2-24
2-14
To avoid equipment damage that could lead to image quality degradation, do not
disconnect a transducer while it is in use. Either freeze the image or switch to
another transducer before disconnecting.
To prevent misdiagnosis or harm to the patient, use the correct transducer for
the application. The diagnostic capability differs for each transducer, exam type,
and imaging mode. Transducers are developed to specific criteria depending on
their physical application. These criteria include biocompatibility requirements.
Understand the system's capabilities prior to use.
"Connecting transducers"