3 Check the battery switches; ensure that all three switches are depressed to the
ON position.
Note
4 Reconnect the system to AC power.
The battery charge indicator at the base of the stand blinks green, and the battery icon on the clinical
monitor displays the battery charging state.
To turn off the system
Note
Press the power button.
The system will power down when your data is safe. Any in-progress transfers will complete when power
is restored.

Adjusting the height and angle

WARNINGS
To raise or lower the platform
While pressing down the height-adjustment pedal, grasp both sides of the platform and push down or
pull up to the desired height.
To lock a wheel
Press down the lever on the wheel.
To unlock the wheel, press up on the bottom of the lever.
To adjust the clinical monitor angle
Grasping the clinical monitor on both sides, tilt or rotate it.
2-4
The system will not charge and maintain the batteries if the battery switches are
depressed to the ·O symbol, which is the OFF position.
If the system appears unresponsive, wait several minutes before restarting it.
Restarting the system while it is performing data-intensive background
activities, such as transferring patient files, can result in loss of patient data. To
power down an unresponsive system, press and hold the power button until the
system shuts down. This procedure may take 5 seconds or longer.
Lock the wheels whenever the system is unattended or stationary.
To avoid possible injury from an unexpected clinical monitor collapse during
system transport, collapse the clinical monitor before system transport (see
collapse the clinical monitor"
on page 2-5).
ʘ
symbol, which is the
"To