The system is intended for use in the electromagnetic environment specified below.
Table 9-3: Manufacturer's Declaration - Electromagnetic Emissions
Emissions Test
Compliance
RF emissions
Group 1
ClSPR 11
RF emissions
Class A
ClSPR 11
Harmonic
Class A
emissions IEC
61000-3-2
Voltage
Complies
fluctuations/
flicker emissions
IEC 61000-3-3
The system is intended for use in the electromagnetic environment specified below.
Table 9-4: Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
Transient burst IEC
61000-4-4
Surge
IEC 61000-4-5
9-16
Electromagnetic Environment
The SonoSite X-Porte ultrasound system uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
The SonoSite X-Porte ultrasound system is suitable for use in all
establishments other than domestic and those directly connected to
the public low-voltage power supply network which supplies
buildings used for domestic purposes.
IEC 60601 Test
Level
± 2.0KV, ± 4.0KV, ±
6.0KV contact
± 2.0KV, ± 4.0KV, ±
8.0KV air
± 2KV on the mains
± 1KV on signal
lines
± 1KV line(s) to
line(s)
± 2KV line(s) to
earth
Compliance Level
± 2.0KV, ± 4.0KV, ±
6.0KV contact
± 2.0KV, ± 4.0KV, ±
8.0KV air
± 2KV on the mains
± 1KV on signal
lines
± 1KV line(s) to
line(s)
± 2KV line(s) to
earth
Electromagnetic Environment
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.