5 Save the calculation: Tap
To calculate Aortic Valve Area (AVA)
The AVA calculation requires a measurement in 2D and two measurements in Doppler. After the
measurements are saved, the result appears in the patient report.
1 In 2D:
a On a frozen 2D image, tap Calcs.
If this control is hidden, tapping More Controls displays it.
b On the AV/LVOT/AI calculation list, tap LVOT D.
c Position the calipers by dragging.
d Save the calculation: Tap
2 In Doppler, measure either LVOT VMax or LVOT VTI.
VMax - In the AV/LVOT/AI calculation list, tap LVOT VMax, position the caliper by dragging, and
then tap
to save the measurement.
VTI - In the AV/LVOT/AI calculation list, tap LVOT VTI, trace the Doppler waveform by dragging, and
then tap
to save the measurement.
3 In Doppler, measure either AV VMax or AV VTI
VMax - In the AV/LVOT/AI calculation list, tap AV VMax, position the caliper by dragging, and then
tap
to save the measurement.
Note
VTI - In the AV/LVOT/AI calculation list, tap AV VTI, trace the Doppler waveform by dragging, and
then tap
to save the measurement.
If VTI is chosen, the VMax value derived from the trace is used as input to the
Notes
If VTI measurements are made for both LVOT and AV, a second AVA result is
To calculate Qp/Qs
The Qp/Qs calculation requires two measurements in 2D and two measurements in Doppler. After the
measurements are saved, the result appears in the patient report.
5-32
under the measurement name.
under the measurement name.
If VTI is chosen, the VMax value derived from the trace is used as input to the AVA
calculation.
AVA calculation.
provided.
Measurements and calculations