FujiFilm SonoSite X-Porte User Manual page 223

Hide thumbs Also See for SonoSite X-Porte:
Table of Contents

Advertisement

a Use either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner
from the list of approved cleaners.
Approved cleaners/disinfectants for the transducer:
Product
b
SaniCloth AF3
SaniCloth Plus
PI-Spray II
a
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a minimum
period of time.
b
Qualified for use as an intermediate-level disinfectant for mycobacteria.
Refer to the cleaners and disinfection document available at
current cleaners and disinfectants.
b Remove all gel, debris, and bodily fluids from the transducer.
c With a new wipe, clean the cable and transducer, starting from the cable, wiping toward the scanhead.
This method helps to avoid cross-contamination.
Caution
d Observe the minimum wet contact time and refer to manufacturer's instructions. Monitor the transducer
for wet appearance. Re-apply with a new wipe if no longer wet.
Cleaning and disinfecting
Compatible transducers
C11xp, C35xp, C60xp, HFL38xp,
HFL50xp, HSL25xp, ICTxp, L25xp,
L38xp, P10xp, P21xp, rP19xp
C60xp, HFL38xp, HFL50xp, ICTxp
C11xp, C60xp, D2xp, HFL38xp,
HFL50xp, HSL25xp, ICTxp, L25xp,
L38xp, P10xp, P21xp, rP19xp
Do not allow moisture near the electronic components of the connector.
Minimum wet contact time
3 minutes
3 minutes
10 minutes
www.sonosite.com
for a complete list of the most
a
8-5