Checking the Air Flow

Thermo Fisher Scientific
Cylinder Cover
Distance Ring

Figure 5-1. Location of Pump Parts

When installing new vanes, be sure that the beveled side points outwards
(see circled areas within
order.

Figure 5-2. Carbon Vane Installation

To check and calibrate the air flow rate, please see the detailed description
given in the "Calibration" chapter. For this purpose, a flow transfer standard
(FTS) is needed to determine and calibrate the air flow (preferred flow
meters can be supplied by Thermo Fisher Scientific as maintenance
accessories).
Figure
5-2). Reassemble the parts in the reverse
Model 5030 SHARP Monitor Instruction Manual
Preventive Maintenance
Checking the Air Flow
Silencer Diaphragm
Silencer
5-3