Quantum Data 802BT User Manual page 576

Video test generator
Hide thumbs Also See for 802BT:
Table of Contents

Advertisement

VSPP
Class
Description
Command syntax
Query syntax
Query response
Other required
commands
Example
566
Appendix A Command Reference
Vertical Sync Pulse Polarity
Format parameter setting - Synchronization
Establishes the logic sense of the digital vertical sync outputs. Setting polarity to 1 causes
the leading edge of vertical sync to be a low-to-high transition. Setting polarity to 0 causes
the leading edge of vertical sync to be a high-to-low transition. The VSPP? query returns
the current polarity of VSPP.
VSPP polarity
polarity
0 = high-to-low transition
1 = low-to-high transition
VSPP?
0 or 1
In order to use digital vertical sync, it must be gated on with the
digital separate H and V sync must be selected with the
command instructs the generator to use the new setting. The
hardware to the new setting and redraws the test image.
VSPP 1
// Set active hi V sync in buffer
VSPG 1
// Enable V sync output in buffer
SSST 1
// Select H&V sync type in buffer
FMTU
// Update hardware to current buffer contents
VSPG
command and
SSST
command. The
FMTU
ALLU
command updates

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

802r

Table of Contents