Temperature
Min and Max Internal
Temperature Limits
Thermo Fisher Scientific
(if installed) indicate that the power supplies are working and connections
are successful. There are no setting screens for these alarms.
In the Main Menu, choose Alarms.
ALARMS:
>INTERNAL TEMP
Internal
The Internal Temperature screen displays the current internal temperature
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 8 to 47 °C. If the internal temperature reading goes beyond
either the minimum or maximum limit, an alarm is activated and the alarm
(bell) icon appears in the status bar on the Run screen and in the Main
Menu.
In the Main Menu, choose Alarms > Internal Temp.

INTERNAL TEMPERATURE:

>MIN
The Minimum Internal Temperature alarm limit screen is used to change
the minimum internal temperature alarm limit. The minimum and
maximum internal temperature screens function the same way.
In the Main Menu, choose Alarms > Internal Temp > Min or Max.
ALARMS DETECTED
BENCH TEMP
PRESSURE
SAMPLE FLOW
RANGE
AVG
DIAGS ALARM
BIAS VOLTAGE
AGC INTENSITY
MOTOR SPEED
ZERO CHECK
SPAN CHECK
ZERO AUTOCAL
SPAN AUTOCAL
CONCENTRATION
O2 CONCENTRATION
EXTERNAL ALARMS
MOTHERBOARD STATUS
INTERFACE STATUS
I/O EXP STATUS
ACTUAL
MAX
RANGE
AVG
DIAGS ALARM
0
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
o
30.1
C
o
8.0
C
o
47.0
C
Model 48i Instruction Manual
Operation
Alarms Menu
3-69