Hold Current Output; Hold Relay Output - Thermo Scientific DensityPRO-T Series User Manual

Gamma density measurement systems

Advertisement

Operation
Commands
Hold Current
Output

Hold Relay Output

-
4
16
DensityPRO User Manual
Figure 4-10. Sample Selection Options for Common Action
The final item on the Common Action dropdown list is Perform Self Test. Currently
this function is nonoperational. It will be utilized in future versions of the EZ Cal II
software.
Below the Common Action dropdown list on the Commands screen are the Hold
Current Output A, B and C dropdown lists. Each of these lists contains the same
options for holding the preferred current.
Do Nothing – No action is taken when the submit button is pressed.
Hold at minimum – The current output value is held at the minimum value
entered.
Hold at maximum – The current output value is held at the maximum value
entered.
Hold at set value – The current output value is held at the set hold value.
Hold at Fault Low – The current output value is held at the Fault Low value.
Hold at Fault High – The current output value is held at the Fault High value.
Clear Hold value – If the Hold/Live value is set to Hold, then it is reset to Live.
The Hold Relay Output A and B sections are for display purposes only and are not
configurable by the user.