Operation
Commands
Hold Current
Output

Hold Relay Output

Alarm Action

-
4
16
LevelPRO User Manual
The final item on the Common Action dropdown list is Perform Self Test. Currently
this function is nonoperational. It will be utilized in future versions of the EZ Cal II
software.
Below the Common Action dropdown list on the Commands screen are the Hold
Current Output A, B and C dropdown lists. Each of these lists contains the same
options for holding the preferred current.
Do Nothing – No action is taken when the submit button is pressed.
Hold at minimum – The current output value is held at the minimum value
entered.
Hold at maximum – The current output value is held at the maximum value
entered.
Hold at set value – The current output value is held at the set hold value.
Hold at Fault Low – The current output value is held at the Fault Low value.
Hold at Fault High – The current output value is held at the Fault High value.
Clear Hold value – If the Hold/Live value is set to Hold, then it is reset to Live.
The Hold Relay Output A and B sections are for display purposes only and are not
configurable by the user.
The Alarm Action dropdown list allows the user to make changes to alarms as a group,
either system-wide or by application.
Do Nothing – No action is taken when sent to the detector.
Clear all alarms for all applications – Currently this option is non-functional and
no action is taken upon its selection. This option will be utilized in future versions
of the EZ Cal II software.
Disable all alarms – Any process alarm or mode/fault alarm that is in an enabled
state is changed to the inhibit state. This option is available for all applications or
individual applications.
Enable all alarms – Any process alarm or mode/fault alarm that is in the inhibit
state is changed to an enabled state. This option is available for all applications or
individual applications.
Erase all process alarm assignments – All configurations associated with setting up
a process alarm are set to the default values. These values include Measurement
ID, Action, Action Delay, Set Point, Clear Point, and the Enable/Disable setting.
This option is available for all applications or individual applications.
Erase all mode/fault alarm assignments – All configurations associated with setting
up a mode/fault alarm are reset to the default value. Enable/Disable setting,
Action. This option is available for all applications or individual applications.
Erase all alarm assignments, all mode/fault/process alarms for all applications –
This option allows the user to perform the Erase all process alarm assignments
action and the Erase all mode/fault alarm assignments action simultaneously. This
option is not available for individual applications.