Receive:
set sp field 1 34 ok
stream per
This command reports the currently set time interval in seconds for
streaming data.
Send:
stream per
Receive:
stream per 10
set stream per number value
number value = | 1 | 2 | 5 | 10 | 20 | 30 | 60 | 90 |120 | 180 | 240 | 300 |
This command sets the time interval between two consecutive streaming
data strings to number value in seconds. The example below sets the
number value to 10 seconds.
Send:
set stream per 10
Receive:
set stream per 10 ok
stream time
This command reports if the streaming data string will have a time stamp
attached to it or not, according to
Send:
stream time
Receive:
stream time 0
set stream time value
This command enables value, where value is to attach or disable time stamp
to streaming data string, according to
attaches a time stamp to streaming data.
Send:
set stream time 0
Receive:
set stream time 0 ok
Table B–5. Stream Time Values
Value
Stream Time
00
Attaches time stamp to streaming data string
01
Disables time stamp to streaming data string
set cal no bkg
set cal nox bkg
set cal pre bkg
These commands will auto-calibrate the NO, NO
backgrounds. If the instrument is set to manual NO
C-Link Protocol Commands
Table
B–5.
Table
B–5. The example below
, and prereactor
x
mode, the response to
x
Model 42i Trace Level Instruction Manual
Calibration
B-25