Appendix C Technical Description: Inlet; Rapid Multi-Stream Sampler (Rms; Introduction; Safety - Thermo Scientific Prima PRO User Manual

Mass spectrometers
Table of Contents

Advertisement

Rapid Multi-
Sampler

Introduction

Thermo Fisher Scientific
Appendix C
Technical Description: Inlet
Stream
The Rapid Multi-Stream Sampler (RMS) inlet system has been used
successfully with a range of Prima and Sentinel gas analysers. The
unit selects a single gas stream from a number of incoming sample or
calibration gas streams, and some of the selected gas stream is then
delivered to the mass spectrometer for analysis.

Safety

Warning! The notes below must be considered in addition to safety
notes elsewhere in this manual covering the system as a whole.
Depending on the nature of the gases in use (sample and calibration
gases), precautions may need to be taken to avoid inhalation of,
contact with, or ignition of these gases. In general, these precautions
are as follows.
Always use the correct tubing size for the fittings supplied and
the correct nut / ferrule assembly.
Do not exceed the pressure and flow specifications detailed
below.
Leak check all external pipe work, and recheck periodically.
Leak check inlet system periodically, as defined in this manual.
Before opening up any gas handling part of the system, it may be
necessary to purge the system with a suitable inert or non-
hazardous gas. Gas lines and the exhaust line should be isolated.
Consideration should be given to the possibility and implications
of failure of the isolation mechanism used.
Warning! It is the responsibility of the instrument user to assess the
potential hazards involved in working on any gas handling parts of
the system.
Prima PRO & Sentinel PRO Mass Spectrometers User Guide
C-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sentinel pro

Table of Contents