7. Instrument Features

7.7 Flash Type Luminescence

STOP
Parameters
Wavelength Range
Measurement Range
Integration Time/well
Cross talk % (black plate)
Attenuation of Light
ATP Detection Limit (Standard PMT)
ATP Detection Limit (Basic PMT)
STOP
122
Switch on the instrument at least 15 minutes before starting a
luminescence measurement. Some components need to warm up to
guarantee stable conditions for the measurement.
Luminescence Detection uses the photon counting technique.
Specifications are only valid when the instrument is placed in a
location with illumination < 500 lux.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
Caution
Characteristics
380 – 600 nm
> 6 orders of magnitude
1 – 20000 ms
< 0.01 %
10, 1 (no attenuation)
< 80 amol/well
< 500 amol/well
Caution
2008-07