Table 11-19
SUBSYSTEM COMMANDS
SENSe[1|2]:CORRection:
COLLect[:ACQuire] <char>
SENSe[1|2]:CORRection
:COLLect:CKIT:PORT[1|2|...|12
][:SELect]<string>
SENSe[1|2]:CORRection:COLLe
ct:CKIT[:SELect]<string>
SCPI Command Summary
SCPI Device Command Summary
SENSe (5 of 16)
command
only
STRING
STRING
11-42
FORM
Measures a calibration standard —
select <char> from:
STANdard1
STANdard2
STANdard3
STANdard4
STANdard5
STANdard6
STANdard7
Selects which port of the analyzer or
multiport test set will be assigned a Cal
Kit. Also, selects a Cal Kit— choose
from one of the following strings:
'COAX,7MM,TYPE-N,50,FEMALE'
'COAX,7MM,TYPE-N,50,MALE'
'COAX,3.5MM,APC-3.5,50,IMPLIED'
'USER|USER1|USER2,IMPLIED,
IMPLIED,IMPLIED, IMPLIED'
'COAX,7MM,TYPE-F,75,IMPLIED'
'COAX,7MM,TYPE-N,75,FEMALE'
'COAX,7MM,TYPE-N,75,MALE'
'COAX,7MM,APC-7,50,IMPLIED'
'COAX,7MM,TYPE-7-16,50,FEMALE'
Selects a Cal Kit— choose from one of
the following strings:
'COAX,7MM,TYPE-N,50,FEMALE'
'COAX,7MM,TYPE-N,50,MALE'
'COAX,3.5MM,APC-3.5,50,IMPLIED'
'USER|USER1|USER2,IMPLIED,
IMPLIED,IMPLIED, IMPLIED'
'COAX,7MM,TYPE-F,75,IMPLIED'
'COAX,7MM,TYPE-N,75,FEMALE'
'COAX,7MM,TYPE-N,75,MALE'
'COAX,7MM,TYPE-7-16,50,FEMALE'
'COAX,7MM,TYPE-7-16,50,MALE'
DESCRIPTION
Programmer's Guide