Introduction - Agilent Technologies 8712ET Programmer's Manual

Rf network analyzers
Hide thumbs Also See for 8712ET:
Table of Contents

Advertisement

NOTE
Using Graphics

Introduction

Introduction
The analyzer has a set of user graphics commands that can be used to
create graphics and messages on the display. The GRAPHICS example
program in the Example Programs Guide uses some of these commands
to draw a simple setup diagram. These commands, listed below, are of
the form:
DISPlay:WINDow[1|2|10]:GRAPhics:<mnemonic>.
The number specified in the WINDow part of the command selects where
the graphics are to be written.
draws the graphics to the channel 1 measurement
WINDow1
screen. (This is the default if no window is specified in
the mnemonic.)
draws the graphics to the channel 2 measurement
WINDow2
screen.
draws the graphics to an IBASIC display partition. The
WINDow10
graphics can only be output in red on a PCL printer.
When graphics commands are used to write directly to a measurement
screen, they write to the static graphics plane (the same plane where the
graticule is drawn). There is no sweep-to-sweep speed penalty once the
graphics have been drawn.
7-2
Programmer's Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8712es8714et8714es