Table 10-11
KEYSTROKES
CAL, More Cal 1, 2
Port Extensions
Port Ext's ON off
Port 1 Extension
(Refl Port Extensions) 3, 5
Port 2 Extension
(Trans Port Extensions)
Velocity Factor
Smith Chart Z0
System Z0
50 Ω
75 Ω
Cal Kit
Type-N(f) (Default)
Type-N(m)
1. The port number in each command is the port selected from the screen using the
and
keys.
down
2. The instrument system impedance determines the set of valid connector choices in
this menu. Connectors with a characteristic impedance other than the instrument
system impedance are dimmed on the screen.
3. This Key is active if port extensions are on.
4. For 8712ES and 8714ES analyzers.
5. For 8712ET and 8714ET analyzers.
Menu Map with SCPI Commands
Menu Map for 8712ET/ES and 8714ET/ES
Functions, 8712ET/ES and 8714ET/ES (1 of 4)
CAL, More CAL
(menu selection only)
(menu selection only)
SENS[1|2]:CORR:EXT [ON|OFF]
3, 4
SENS[1|2]:CORR:EXT:REFL:TIME <num> s
3, 4
SENS[1|2]:CORR:EXT:TRAN:TIME <num> s
3, 5
SENS[1|2]:CORR:RVEL:COAX <num>
SENS[1|2]:CORR:IMP:INP:MAGN <num> OHM
SENS[1|2]:CORR:IMP:INP:MAGN:SEL
Z0_[50|75]
SENS[1|2]:CORR:EXT:REFL:TIME <num> s
SENS[1|2]:CORR:EXT:REFL:TIME <num> s
(menu selection only)
SENS:CORR:COLL:CKIT:PORT[1|2]
'COAX,7MM,TYPE-N,75,FEMALE'
SENS:CORR:COLL:CKIT:PORT[1|2]
'COAX,7MM,TYPE-N,75,MALE'