Agilent Technologies 8712ET Programmer's Manual page 158

Rf network analyzers
Hide thumbs Also See for 8712ET:
Table of Contents

Advertisement

Table 10-11
KEYSTROKES
1
,
CAL
Test Set Cal
Create "TSET_CAL"
XX Ports
Enh Resp 1 - Port 3
2 - Port
Measure Opens
Measure Shorts
Measure Loads
Measure Thrus
All Stds Done
Periodic SelfCal
SelfCal Once
SelfCal Timer
SelfCal All Ports
2-Port Cal on OFF
Isolation on OFF
1. For use with multiport test sets only.
2. You select the number of test set ports, and a guided calibration procedure begins.
3. For 8712ES and 8714ES analyzers only.
Menu Map with SCPI Commands
Menu Map for 8712ET/ES and 8714ET/ES
,
Functions, 8712ET/ES and 8714ET/ES
CAL
Test Set Cal
2
3
3
3
10-30
SCPI COMMAND
SENS[1|2]:CORR:TESTSET;*WAI
SENS[1|2]:CORR:COLL:METHOD TEST;
SENS[1|2]:CORR:COLL:PORTS
<2|4|6|8|10|12>
SENS[1|2]:CORR:COLL:METHOD TEST 1
SENS[1|2]:CORR:COLL:METHOD TEST 2
SENS[1|2]:CORR:COLL:MP:OPEN
<STAN1|STAN2|...|STAN12>;*WAI;
SENS[1|2]:CORR:COLL:MP:SHORT
<STAN1|STAN2|...|STAN12>;*WAI;
SENS[1|2]:CORR:COLL:MP:LOAD
<STAN1|STAN2|...|STAN12>;*WAI;
SENS[1|2]:CORR:COLL:MP:THRU
<STAN1|STAN2|...|STAN6>;*WAI;
SENS[1|2]:CORR:COLL:SAVE;*WAI;
CAL:SELF ON
CAL:SELF ONCE
CAL:SELF:TIMER <num>
CAL:SEL0F:ALL
CAL:SELF:METHOD [ONEP|TWOP]*WAI
SENS[1|2]:CORR:ISOL [ON|OFF]*WAI
Programmer's Guide