Agilent Technologies 8712ET Programmer's Manual page 145

Rf network analyzers
Hide thumbs Also See for 8712ET:
Table of Contents

Advertisement

Table 10-4
KEYSTROKES
, (Transmissn) (continued)
CAL
Default Response
Default 2-Port
User Response
Default Response
Response
Measure Standard
Response & Isolation
Measure Standard 2
Enhanced Response
Measure Standard
1. Pressing this key begins a guided calibration procedure using one standard. When
the calibration is complete, the command
*WAI;:SENS[1|2]:CORR:COLL:SAVE;*WAI is executed.
2. Pressing this key begins a guided calibration procedure using two standards. When
the calibration is complete, the command
*WAI;:SENS[1|2]:CORR:COLL:SAVE;*WAI is executed.
3. Pressing this key begins a guided calibration procedure using four standards.
When the calibration is complete, the command *WAI;:SENS[1|2]:CORR:COLL:
SAVE;*WAI is executed. The order of appearance and the content of the softkey
menus depend on the measurement channel 1 and measurement channel 2 configu-
ration.
, Functions, 8712ES and 8714ES (2 of 3)
CAL
SENS[1|2]:CORR:CSET DEF; *WAI
SENS[1|2]:CORR:CLASS DEF2; *WAI
SENS[1|2]:CORR:CLASS DEF1; *WAI
SENS1:CORR:CSET DEF; *WAI
SENS[1|2:CORR:COLL:IST OFF;METH TRAN1; *WAI
1
SENS[1|2]:CORR:COLL STAN[1|2|3];*WAI
SENS[1|2:CORR:COLL:IST OFF;METH TRAN2; *WAI
SENS[1|2]:CORR:COLL STAN[1|2|3];*WAI
SENS[1|2:CORR:COLL:IST OFF;METH TRAN3; *WAI
3
SENS[1|2]:CORR:COLL STAN[1|2|3];*WAI
Programmer's Guide
Menu Map with SCPI Commands
Menu Map for 8712ET/ES and 8714ET/ES
SCPI COMMAND
10-17