SCPI Error Messages
Execution Errors
Error
Number
−220
Parameter error — Indicates that a program data element related error
occurred. This error message should be used when the device cannot detect
the more specific errors −221 through −229.
−221
Settings conflict — Indicates that a legal program data element was parsed
but could not be executed due to the current device state.
−222
Data out of range — Indicates that a legal program data element was
parsed but could not be executed because the interpreted value was outside
the legal range as defined by the device.
−223
Too much data — Indicates that a legal program data element of block,
expression, or string type was received that contained more data than the
device could handle due to memory or related device-specific requirements.
−224
Illegal parameter value — Used where an exact value, from a list of
possible values, was expected.
−225
Out of memory — The device has insufficient memory to perform the
requested operation.
−226
Lists not same length — Attempted to use LIST structure having
individual LIST's of unequal lengths.
−230
Data corrupt or stale — Possibly invalid data; new reading started but not
completed since last access.
−231
Data questionable — Indicates that measurement accuracy is suspect.
−232
Invalid format — Indicates that a legal program data element was parsed
but could not be executed because the data format or structure is
inappropriate, such as when loading memory tables or when sending a
SYSTem:SET parameter from an unknown instrument.
−233
Invalid version — Indicates that a legal program data element was parsed
but could not be executed because the version of the data is incorrect to the
device. This particular error should be used when file or block data formats
are recognized by the instrument but cannot be executed for reasons of
version incompatibility. For example, an unsupported file version, or an
unsupported instrument version.