Table 11-19
SUBSYSTEM COMMANDS
SENSe[1|2]:CORRection:COLL
ect:ISTate[:AUTO]<ON|OFF>
SENSe[1|2]:CORRection:
COLLect:METHod <char>
SENSe[1|2]:CORRection:COLL
ect:MP:OPEN
<STAN1|STAN2|...|STAN12>
SENSe[1|2]:CORRection:COLL
ect:MP:SHORT
<STAN1|STAN2|...|STAN12>
1. Binary parameters accept the values of 1(on) and 0(off) in addition to ON and OFF.
2. For use with multiport test sets only.
SENSe (6 of 16)
FORM
NR1
1
command
only
command
only
command
only
Programmer's Guide
SCPI Device Command Summary
DESCRIPTION
Selects the instrument state for
calibration — choose Full Band (ON) or
User Defined (OFF).
Selects the type of calibration — choose
from:
TRAN1 — Transmission response
TRAN2 — Transmission response &
Isolation
TRAN3 — Transmission enhanced
response
REFL3 — Reflection one port
TESTset — Test Set Calibration
(Enhanced Response/1-Port)
TESTset1 — Test Set Calibration
(Enhanced Response/1-Port)
TESTset2 — Test Set Calibration
1
(2-Port)
VERIFY — Calibration Check
TWOPort — Two-port calibration
NONE — No calibration
Measures an open on the port selected
during a test set calibration.
Measures a short on the port selected
during a test set calibration.
SCPI Command Summary
1
1
2
2
11-43