Agilent Technologies 8712ET Programmer's Manual page 147

Rf network analyzers
Hide thumbs Also See for 8712ET:
Table of Contents

Advertisement

Table 10-5
KEYSTROKES
(Reflection)
CAL,
Default 1-Port
1-Port
Measure Standard
Measure Standard — Open
Measure Standard — Short
Measure Standard — Load
Normalize on OFF
Cal Check
(see
Table 10-10 on page
1. Pressing this softkey begins a guided 1-port calibration procedure, using three
standards. When the calibration is complete, the command *WAI;:SENS[1|2]
:CORR:COLL:SAVE;*WAI is executed. The order of appearance and the content of
the softkey menus depend on the measurement channel 1 and measurement
channel 2.
Functions, 8712ET and 8714ET (1 of 3)
CAL
(hardkey entry)
SENS[1|2]:CORR:CSET DEF; *WAI
SENS[1|2]:CORR:COLL:IST OFF;METH REFL3;
*WAI
1
SENS[1|2]:CORR:COLL:IST OFF;METH REFL3;
*WAI
SENS[1|2]:CORR:COLL STAN1;*WAI
SENS[1|2]:CORR:COLL STAN2;*WAI
SENS[1|2]:CORR:COLLSTAN3;*WAI;
:SENS[1|2]:CORR:COLL:SAVE;*WAI
TRAC CH[1|2]SMEM,CH[1|2]SDATA;:CALC[1|2]:
MATH (IMPL/CH[1|2]SMEM);:DISP:WIND[1|2]:
TRAC[1|2] ON;TRAC[1|2] OFF
(menu selection only)
10-25)
Programmer's Guide
Menu Map with SCPI Commands
Menu Map for 8712ET/ES and 8714ET/ES
SCPI COMMAND
10-19