13.2.4
To Set Clock Recovery
For setting methods of the clock recovery in the jitter, refer to descriptions in "
Clock Recovery
NOTE
When the measurement item is "TIE", you need to set the clock recovery. For other
measurement items, you do not need to set the clock recovery.
13.2.5
To Set the Jitter Measurement
Press Measure to enter the jitter measurement menu. You can set the jitter
measurement item and the graph display mode of the jitter measurement results.
•
Set the Jitter Measurement Item
The jitter measurement items supported by this oscilloscope include TIE, Cycle to
Cycle, +Width to +Width, and –Width to -Width.
Press Item, then rotate the multifunction knob
measurement item, and then press down the knob to select the item. You can
also press Item continuously or enable the touch screen to select it.
•
TIE
Compares edges in a signal under test with the edges in an ideal signal
determined by the clock recovery feature. Then, all the signal intervals are
measured based on the ideal data rate to generate the error statistics.
•
Cycle to Cycle
Measures the first cycle and then subtracts the first cycle from the second cycle.
The measurement is made according to this rule.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Real-time Eye Analysis
" in the "
Real-time Eye Diagram and Jitter Analysis
" section.
to select desired
(Option)
To Set
219