Calculate:ddemod:marker<N>:Y[:Real]; Calculate:marker<N>[:Set]:Center - Rigol SA5000 Series Programming Manual

Vector signal analysis
Table of Contents

Advertisement

RIGOL
Return Format
The query returns the marker's quadrature component (imaginary) of the Y value in scientific notation.
Example
The following command sets the quadrature component Y value of Marker 1 to 0.435.
:CALCulate:DDEMod:MARKer1:Y:IMAGinary 0.435
The following query returns 4.350000000e-01.
:CALCulate:DDEMod:MARKer1:Y:IMAGinary?
:CALCulate:DDEMod:MARKer<n>:Y[:REAL]
Syntax
:CALCulate:DDEMod:MARKer<n>:Y[:REAL] <real>
:CALCulate:DDEMod:MARKer<n>:Y[:REAL]?
Description
Sets the Y-axis value of the specified marker.
Queries the Y-axis value of the specified marker.
Parameter
Name
<n>
<real>
Consecutive Real Number
Remarks
This command sets the marker's in-phase component (real part) of the Y value.
Return Format
The query returns the Y-axis value of the marker or the in-phase component of the Y value in scientific
notation.
Example
The following command sets the Y-axis value of Marker 1 to 0.325.
:CALCulate:DDEMod:MARKer1:Y:REAL 0.325
The following query returns 3.250000000e-01.
:CALCulate:DDEMod:MARKer1:Y:REAL?
:CALCulate:MARKer<n>[:SET]:CENTer
Syntax
:CALCulate:MARKer<n>[:SET]:CENTer
Description
Sets the center frequency of the analyzer to the frequency of the specified marker.
Parameter
Name
<n>
Remarks
If the marker mode of the specified marker is Position or Fixed, the center frequency will be set to the
frequency of the marker.
2-12
Type
Discrete
Type
Discrete
Range
1|2|3|4|5|6|7|8
--
Range
1|2|3|4|5|6|7|8
Chapter 2 Command System
Default
--
--
Default
--
VSA Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rsa5065Rsa5032Rsa5065-tgRsa5032-tg

Table of Contents