Rigol MSO8064 User Manual

Rigol MSO8064 User Manual

Digital oscilloscope
Hide thumbs Also See for MSO8064:
Table of Contents

Advertisement

Quick Links

User Guide
MSO8000 Series Digital
Oscilloscope
Mar. 2021
.
RIGOL TECHNOLOGIES CO., LTD

Advertisement

Table of Contents
loading

Summary of Contents for Rigol MSO8064

  • Page 1 User Guide MSO8000 Series Digital Oscilloscope Mar. 2021 RIGOL TECHNOLOGIES CO., LTD...
  • Page 3: Guaranty And Declaration

    Notices RIGOL products are covered by P.R.C. and foreign patents, issued and pending. ⚫ RIGOL reserves the right to modify or change parts of or all the specifications ⚫ and pricing policies at the company’s sole decision. Information in this publication replaces all previously released materials.
  • Page 4: Safety Requirement

    RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the instrument properly.
  • Page 5 Do Not Operate with Suspected Failures. If you suspect that any damage may occur to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
  • Page 6: Safety Notices And Symbols

    RIGOL Safety Notices and Symbols Safety Notices in this Manual: WARNING Indicates a potentially hazardous situation or practice which, if not avoided, will result in serious injury or death. CAUTION Indicates a potentially hazardous situation or practice which, if not avoided, could result in damage to the product or loss of important data.
  • Page 7: Measurement Category

    RIGOL Measurement Category Measurement Category MSO8000 series digital oscilloscopes can make measurements in Measurement Category I. WARNING This oscilloscope can only be used for measurements within its specified measurement categories. Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS.
  • Page 8: Working Environment

    RIGOL Working Environment Temperature Operating: 0℃ to +50℃ Non-operating: -30℃ to +70℃ Humidity Operating: Below +30℃: ≤90%RH (without condensation) +30℃ to +40℃: ≤75%RH (without condensation) +40℃ to +50℃: ≤45%RH (without condensation) Non-operating: Below 65℃: ≤90% RH (without condensation) WARNING To avoid short circuit inside the instrument or electric shock, never operate the instrument in a humid environment.
  • Page 9: Care And Cleaning

    RIGOL Pollution Degree Definition Pollution Degree 1: No pollution or only dry, nonconductive pollution occurs. The pollution has no effect. For example, a clean room or air-conditioned office environment. Pollution Degree 2: Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected. For example, indoor environment.
  • Page 10: Environmental Considerations

    RIGOL Environmental Considerations The following symbol indicates that this product complies with the WEEE Directive 2002/96/EC. Product End-of-Life Handling The equipment may contain substances that could be harmful to the environment or human health. To avoid the release of such substances into the environment and avoid harm to human health, we recommend you to recycle this product appropriately to ensure that most materials are reused or recycled properly.
  • Page 11: Mso8000 Series Overview

    MSO8000 Series Overview MSO8000 series is a medium and high-end mixed signal digital oscilloscope designed on the basis of the ASIC chip (RIGOL self-owns its intellectual property right) and UltraVision II technical platform developed by RIGOL. Integrating 7 independent instruments into one, the MSO8000 series is equipped with the analog bandwidth of up to 2 GHz, extremely high memory depth of 500 Mpts, sound waveform display effects, excellent waveform capture rate, and powerful data analysis functions.
  • Page 12: Document Overview

    RIGOL Document Overview Main Topics of this Manual: Chapter 1 Quick Start Introduces the preparations before using the oscilloscope and provides a basic introduction of the instrument. Chapter 2 To Set the Vertical System Introduces the vertical system functions of the oscilloscope.
  • Page 13 RIGOL Chapter 13 Reference Waveform Introduces how to compare the input waveform with the reference waveform. Chapter 14 Pass/Fail Test Introduces how to monitor the input signal by using the Pass/Fail test. Chapter 15 Waveform Recording & Playing Introduces the waveform recording & playing function.
  • Page 14 No. of Model Analog Analog Waveform Generator Digital Bandwidth Channels Channels Channels 16 (Required MSO8064 600 MHz 2 (Opt.) to purchase the probe) 16 (Required MSO8104 1 GHz 2 (Opt.) to purchase the probe) 16 (Required MSO8204 2 GHz 2 (Opt.)
  • Page 15 RIGOL Manuals of this Product: Quick Guide, User Guide, Programming Guide, Data sheet, etc. For the latest version of this manual, download it from the official website of RIGOL (www.rigol.com). MSO8000 User Guide XIII...
  • Page 16: Table Of Contents

    Contents RIGOL Contents Guaranty and Declaration ................. I Safety Requirement ................. II General Safety Summary................II Safety Notices and Symbols ..............IV Measurement Category ................V Ventilation Requirement ................V Working Environment ................VI Care and Cleaning .................. VII Environmental Considerations ..............VIII MSO8000 Series Overview ..............
  • Page 17 Contents RIGOL Drag ..................... 1-27 Rectangle Drawing ................. 1-27 Parameter Setting Method ..............1-29 To Use the Security Lock ............... 1-30 To Use the Built-in Help System ............. 1-31 To View the Option Information and the Option Installation ...... 1-32 Chapter 2 To Set the Vertical System ..........
  • Page 18 Contents RIGOL Chapter 5 To Trigger the Oscilloscope ..........5-1 Trigger Source ..................5-2 Trigger LEVEL/Threshold Level ..............5-3 Trigger Mode ..................5-4 Trigger Coupling..................5-6 Trigger Holdoff ..................5-6 Noise Rejection ..................5-7 Trigger Type ..................5-8 Edge Trigger ................... 5-9 Pulse Trigger .................
  • Page 19 Contents RIGOL Ln ....................6-19 Exp ....................6-20 Abs ....................6-20 Low Pass ..................6-21 High Pass ..................6-22 Band Pass ..................6-23 Band Stop ..................6-24 AX+B .................... 6-26 Avg ....................6-27 Math Operation Label ..............6-28 Auto Measurement ................6-29 Quick Measurement after AUTO............
  • Page 20 Contents RIGOL To Set the Histogram Range ..............9-3 To Enable or Disable the Statistical Function ..........9-4 To Reset ....................9-4 Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) ..10-1 Real-time Eye Analysis ................10-2 To Enable or Disable the Eye Analysis Function ......... 10-2 To Select the Eye Source ..............
  • Page 21 Contents RIGOL To Clear the Display of the Reference Waveform ........13-3 To View Details of the Reference Waveform ..........13-3 To Reset the Reference Waveform ............13-4 Color Setting ..................13-4 Label Setting ..................13-4 To Export to Internal or External Memory ..........13-5 To Import from Internal or External Memory ...........
  • Page 22 Contents RIGOL Exp.Fall..................18-7 ECG ..................... 18-7 Gauss ................... 18-7 Lorentz ..................18-8 Haversine ..................18-8 To Output the Arbitrary Waveform ............18-9 To Load the Channel and Waveform ..........18-9 To Create the Waveform ............... 18-10 To Edit Waveforms ............... 18-12 Modulation ..................
  • Page 23 Contents RIGOL Ref Clock ..................20-9 Help ..................... 20-9 Self-calibration ................20-10 Auto Config ................. 20-11 Print Setting ................20-11 Email ..................20-12 Key Locker .................. 20-13 Quick Operation ................20-14 Screen Saver ................20-17 Self-check ................... 20-18 System Time ................20-19 Default Option ................
  • Page 25: Chapter 1 Quick Start

    Chapter 1 Quick Start RIGOL Chapter 1 Quick Start This chapter introduces the precautions when using the oscilloscope for the first time, the front/rear panels of the oscilloscope, the user interface, touch screen controls, and how to use the built-in help system.
  • Page 26: General Inspection

    The consigner or carrier shall be liable for the damage to the instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the instrument. 2. Inspect the instrument In case of any mechanical damage, missing parts, or failure in passing the electrical and mechanical tests, contact your RIGOL sales representative.
  • Page 27: Appearance And Dimensions

    Chapter 1 Quick Start RIGOL Appearance and Dimensions Figure 1-1 Front View Unit: mm Figure 1-2 Vertical View Unit: mm MSO8000 User Guide...
  • Page 28: To Prepare For Use

    Chapter 1 Quick Start RIGOL To Prepare for Use To Adjust the Supporting Legs Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as better operation and observation.
  • Page 29: Turn-On Checkout

    Chapter 1 Quick Start RIGOL Turn-on Checkout When the oscilloscope is connected to power, press the Power key at the lower-left corner of the front panel to start the oscilloscope. (You can also press Utility → System → Power status, and select "Switch On". After the instrument is connected to power source, it will start directly.
  • Page 30: To Connect The Probe

    To Connect the Probe RIGOL provides the passive probe, the active probe, and the logic probe for MSO8000 Series MSO8000 series. For specific probe models, please refer to Datasheet .
  • Page 31 Figure 1-9. 2. Connect the other terminal of the logic probe to the signal terminal under test. RIGOL's MSO8000 series has an optional configuration of a logic probe RPL2316. To apply to different application scenarios, RPL2316 provides three connection...
  • Page 32 Chapter 1 Quick Start RIGOL RPL2316 Logic methods to connect the signal under test. For details, refer to Probe User Guide Figure 1-9 To Connect the Logic Probe Note: Before you connect the logic probe to the signal under test, connect logic ⚫...
  • Page 33: Function Inspection

    Chapter 1 Quick Start RIGOL Function Inspection Press Default on the front panel, then a prompt message "Restore default settings?" is displayed. Press OK or tap OK to restore the instrument to its factory default settings. Connect the ground alligator clip of the probe to the "Ground Terminal" as shown in Figure 1-10 below.
  • Page 34: Probe Compensation

    Chapter 1 Quick Start RIGOL The signal output from the probe compensation connector can only be used for probe compensation adjustment and cannot be used for calibration. Probe Compensation When the probes are used for the first time, you should compensate the probes to make them match the input channels of the oscilloscope.
  • Page 35: Front Panel Overview

    Chapter 1 Quick Start RIGOL Front Panel Overview 3 4 5 6 7 8 9 10 11 12 13 Figure 1-13 Front Panel Table 1-1 Front Panel Description Description Description Capacitive Touch Screen RUN/STOP Key Function Menu Operation Single Trigger Control Key...
  • Page 36: Rear Panel Overview

    Chapter 1 Quick Start RIGOL Rear Panel Overview Figure 1-14 Rear Panel 1. Handle Rotate the handle upright to carry the instrument easily. Rotate it downward if you do not need to carry it. 2. LAN Interface Connect the instrument to network via this interface. The instrument is in compliance with the standards specified in LXI Device Specification 2011.
  • Page 37 Press Online upgrade or enable the touch screen to tap "Online upgrade", then a "System Update Information" window is displayed, requesting you whether to accept or cancel "RIGOL PRODUCT ONLINE UPGRADE SERVICE TERMS". Tap "Accept" to start online upgrade. Tap "Cancel"...
  • Page 38 Chapter 1 Quick Start RIGOL 6. 10 M In/Out Indicates the reference clock. It can provide accurate sample clock signal for the oscilloscope, and can be used to synchronize the clock of two or multiple oscilloscopes. 7. EXT TRIG Inputs the external signal via the external input terminal.
  • Page 39: Front Panel Function Overview

    Chapter 1 Quick Start RIGOL Front Panel Function Overview Vertical 1, 2, 3, 4: indicates the analog channel switch key. The four channels are ⚫ marked by different colors which are also used to mark both the corresponding waveforms of the specified channel on the screen and the channel input connectors.
  • Page 40: Horizontal

    Chapter 1 Quick Start RIGOL scale, and turn it counterclockwise to increase the scale. During the modification, the display amplitude of the waveform will enlarge or reduce. The scale information in the corresponding status label will change accordingly. Press down this knob to quickly switch the vertical scale adjustment mode between "Coarse"...
  • Page 41: Wave

    Chapter 1 Quick Start RIGOL of the screen would change accordingly. Press down this knob to quickly switch the horizontal time base adjustment mode between "Coarse" and "Fine". Wave Math: indicates the math operation key. Press this key to enable the math ⚫...
  • Page 42: Trigger

    Chapter 1 Quick Start RIGOL Wave Vertical SCALE: indicates the waveform vertical scale knob. Rotate ⚫ the knob to adjust the vertical scale of math operation waveform or the vertical scale of the reference waveform, and also can adjust the size of the displayed digital channel waveforms.
  • Page 43: Run/Stop

    Chapter 1 Quick Start RIGOL RUN/STOP Press this key to set the operating state of the oscilloscope to "RUN" or "STOP". In the "RUN" state, the key is illuminated in yellow. In the "STOP" state, the key is illuminated in red.
  • Page 44: Function Menu

    Chapter 1 Quick Start RIGOL Function Menu Measure: Press this key to enter the measurement setting menu. You can set the measurement source, enable or disable the all measurement function, the statistical analysis function, and etc. You can make a quick measurement for 41 waveform parameters.
  • Page 45: Quick Key (Shortcut Key)

    Chapter 1 Quick Start RIGOL Quick Key (Shortcut Key) Press this key to perform the quick operation, including screenshot, waveform saving, setup saving, all measurement, and statistics reset. Press Utility → More → Quick settings to set the quick shortcut keys.
  • Page 46: User Interface

    Chapter 1 Quick Start RIGOL User Interface MSO8000 series has a 10.1-inch WSVGA (1024× 600) LCD, with 256 level gray-scale display. The user interface displays the acquired waveforms, the setting information, and the measurement results. 5 6 7 10 11 12...
  • Page 47 Chapter 1 Quick Start RIGOL 4. Sample Rate/Memory Depth Displays the current sample rate and memory depth of the analog channel. ⚫ The sample rate and the memory depth will change along with the ⚫ horizontal time base. 5. Auto Measurement Label Enable the touch screen, and then tap the label.
  • Page 48 Chapter 1 Quick Start RIGOL The trigger level label is displayed at the left section of the screen, and ⚫ the trigger level/threshold level value is displayed at the upper-right corner of the screen. When you use Trigger LEVEL to modify the trigger level/threshold ⚫...
  • Page 49 Chapter 1 Quick Start RIGOL 18. CH4 Status Label Displays the status of CH4. ⚫ Displays the vertical scale of CH4. That is, the voltage value per grid of CH4 ⚫ in the vertical axis. Displays the offset of CH4, i.g. the vertical offset of CH4 waveform.
  • Page 50: Touch Screen Controls

    Chapter 1 Quick Start RIGOL digital channel is consistent with that of the channel label, being displayed in red. The grouping setting function under the logic analyzer function menu can divide the digital channels into 4 channel groups. The channel label of the same channel group is displayed in the same color;...
  • Page 51: Pinch & Stretch

    Chapter 1 Quick Start RIGOL Pinch & Stretch Pinch or stretch two points on the screen with two fingers to zoom in or out the waveform. To zoom in the waveform, first pinch the two fingers and then stretch the fingers;...
  • Page 52 Chapter 1 Quick Start RIGOL you can tap to select "Trigger zone A", "Trigger zone B", "Histogram", "Horizontal zoom in", "Vertical zoom in", or "Waveform zoom in". Drag a finger from lower right to upper left across the screen to draw a rectangle on the screen, as shown in Figure 1-19 (b).
  • Page 53: Parameter Setting Method

    Chapter 1 Quick Start RIGOL Parameter Setting Method You can use the knob or enable the touch screen to set the parameters of MSO8000 series. The common parameter setting methods are as follows: Method 1: For the parameters with the sign...
  • Page 54: To Use The Security Lock

    Chapter 1 Quick Start RIGOL Method 4: For the parameters without the above signs, press the desired menu softkey to switch between the parameter items. This method is applicable to the parameters with only two available options. The above method is commonly used for the parameter settings of the oscilloscope.
  • Page 55: To Use The Built-In Help System

    Chapter 1 Quick Start RIGOL To Use the Built-in Help System The help system of this oscilloscope provides instructions for all the function keys on the front panel and their corresponding menu keys. The steps for opening the built-in help system are as follows: Press Utility →...
  • Page 56: To View The Option Information And The Option Installation

    Chapter 1 Quick Start RIGOL Method 3: Enable the touch screen and tap the desired help option to get the corresponding help information. Help information for other keys and buttons: ⚫ Power key: powers on/off the instrument. ➢ Menu Off key: displays or hides menus. Pressing this key can hide the ➢...
  • Page 57 Then, you can install the option according to the following steps. 1) Obtain an option license (1) Log in to the RIGOL official website (www.rigol.com), click License Activation to enter the "Registered product license code" interface. (2) In the software license registration interface, input the correct key, serial number (press Utility →...
  • Page 58 Chapter 1 Quick Start RIGOL Only 1 option license file of one instrument is allowed to be stored in the ⚫ same USB storage device, but the USB storage device can store the option license file of several different instruments. You are not allowed to modify the licensed filename.
  • Page 59: Chapter 2 To Set The Vertical System

    Chapter 2 To Set the Vertical System RIGOL Chapter 2 To Set the Vertical System MSO8000 series provides four analog input channels (CH1-CH4), and each channel is equipped with an independent vertical control system. The setting methods for the vertical systems of the four channels are the same. This chapter takes CH1 as an example to introduce the setting method for the vertical system.
  • Page 60: To Enable Or Disable The Analog Channel

    Chapter 2 To Set the Vertical System RIGOL To Enable or Disable the Analog Channel Enable the analog channel: Connect a signal to the channel connector of CH1, and then press 1 in the vertical control area (Vertical) on the front panel to enable CH1. Then, the backlight of the channel is illuminated.
  • Page 61: To Adjust The Vertical Scale

    Chapter 2 To Set the Vertical System RIGOL To Adjust the Vertical Scale Vertical scale indicates the voltage value per grid in the vertical axis of the screen. It is often expressed in V/div. While you adjust the vertical scale, the display amplitude of the waveform would enlarge or reduce.
  • Page 62: Vertical Expansion

    Chapter 2 To Set the Vertical System RIGOL Coarse adjustment (take counterclockwise as an example): Rotate Vertical ⚫ SCALE to set the vertical scale at 1-2-5 step, i.e. 1 mV/div, 2 mV/div, 5 mV/div, 10 mV/div…10 V/div. Vertical Expansion When changing the vertical scale of the analog channel by rotating the Vertical SCALE knob, you can select to expand or compress the waveform around the "Center"...
  • Page 63: Channel Coupling

    Chapter 2 To Set the Vertical System RIGOL of the screen. The following window is displayed. Tap the icon at the right side of the Offset input field to decrease or increase the offset. You can also tap the Offset input field to input a specific value with the displayed numeric keypad.
  • Page 64: Bandwidth Limit

    Chapter 2 To Set the Vertical System RIGOL Bandwidth Limit MSO8000 series supports the bandwidth limit function. Setting the bandwidth limit can reduce the noises in the displayed waveforms. For example, the signal under test is a pulse with high frequency oscillation.
  • Page 65: Input Impedance

    Chapter 2 To Set the Vertical System RIGOL Table 2-2 Probe Ratio Attenuation Ratio Menu (display amplitude of the signal: actual amplitude of the signal) 0.0001X 0.0001:1 0.0002X 0.0002:1 0.0005X 0.0005:1 0.001X 0.001:1 0.002X 0.002:1 0.005X 0.005:1 0.01X 0.01:1 0.02X 0.02:1...
  • Page 66: Waveform Invert

    Chapter 2 To Set the Vertical System RIGOL Press 1 to open the setting menu of CH1. Then, press Impedance continuously to select the input impedance of the oscilloscope. If you select "50 Ω", the channel status label at the bottom of the screen will display the icon...
  • Page 67: Passive Probe

    If not recognized, please refer to "Probe Ratio" to set the probe ratio. Active Probe If an active probe is connected, for example, when RIGOL's RP7150 probe model is connected to the oscilloscope, press Probe to open the probe menu, and you can operate on the menu item.
  • Page 68: Amplitude Unit

    Chapter 2 To Set the Vertical System RIGOL or use the numeric keypad to set the DC bias voltage. Probe Information ⚫ Press About to view information about the probe currently connected, such as the manufacturer, model, serial number, and the last calibration date.
  • Page 69: To Set The Bias

    Chapter 2 To Set the Vertical System RIGOL numeric keypad to set the desired delay calibration time. The available range of the delay calibration time is from -100 ns to 100 ns. Note: This parameter is related to the instrument model and the current horizontal time base setting.
  • Page 70 Chapter 2 To Set the Vertical System RIGOL "To Create a Folder". Label Input Area Input Selection Area Virtual Keypad Delete Key Upper/Lower Case Switch (enabled in English Input Mode) Figure 2-3 Label Editing Interface For example, set the label to "...
  • Page 71 Chapter 2 To Set the Vertical System RIGOL You can also enable the touch screen to drag the channel label to move its position on the screen. You can run the :CHANnel<n>:LABel:POSition command to set the position of the specified channel label; run the :CHANnel<n>:LABel:SIZe command to set the font size of the specified channel label.
  • Page 73: Chapter 3 To Set The Horizontal System

    Chapter 3 To Set the Horizontal System RIGOL Chapter 3 To Set the Horizontal System Contents in this chapter: To Adjust the Horizontal Time Base ◼ To Adjust the Horizontal Position ◼ Delayed Sweep ◼ MSO8000 User Guide...
  • Page 74: To Adjust The Horizontal Time Base

    Chapter 3 To Set the Horizontal System RIGOL To Adjust the Horizontal Time Base Horizontal time base, also called the horizontal scale, refers to the time of each grid in the horizontal direction of the screen. It is usually expressed in s/div. The adjustable range of the horizontal time base is related to the model type.
  • Page 75: To Adjust The Horizontal Position

    Chapter 3 To Set the Horizontal System RIGOL within the adjustable range. Rotate it clockwise to reduce the horizontal time base, and counterclockwise to increase the horizontal time base. Note: When the delayed sweep is enabled, rotate the Horizontal SCALE knob to adjust the time base for the delayed sweep.
  • Page 76: Delayed Sweep

    Chapter 3 To Set the Horizontal System RIGOL Note: When the delayed sweep is enabled, rotate the Horizontal POSITION knob to adjust the horizontal position of the delayed sweep (press down the knob to quickly reset it). The horizontal position of the delayed sweep at the upper-right corner of the screen changes accordingly.
  • Page 77 Chapter 3 To Set the Horizontal System RIGOL Waveform before enlargement: The waveform in the area that is not covered by subtransparent blue in the upper part of the screen is the waveform before enlargement. Its horizontal time base (also called main time base) is displayed at the upper-left corner of the screen.
  • Page 79: Chapter 4 To Set The Sample System

    Chapter 4 To Set the Sample System RIGOL Chapter 4 To Set the Sample System Contents in this chapter: Timebase Mode ◼ Acquisition Mode ◼ Sampling Mode ◼ Sample Rate ◼ LA Sample Rate ◼ Memory Depth ◼ LA Memory Depth ◼...
  • Page 80: Timebase Mode

    Chapter 4 To Set the Sample System RIGOL Timebase Mode MSO8000 series supports three available time base modes: YT mode, XY mode, and ROLL mode. By default, the time base mode is YT. Press Acquire → Timebase Mode on the front panel, and then rotate the multifunction knob to select the desired time base mode.
  • Page 81 Chapter 4 To Set the Sample System RIGOL According to sin=A/B or C/D,  is the phase deviation angle between the two channels. The definitions of A, B, C, and D are shown in the above figure. The phase deviation angle is obtained, that is: =arcsin(A/B) or arcsin(C/D)
  • Page 82: Roll Mode

    Chapter 4 To Set the Sample System RIGOL Note: The maximum sample rate in XY mode is 2.5 GSa/s. In this mode, CH1 and CH2 ⚫ are forced to be enabled; CH3 and CH4 are forced to be disabled. Generally, a longer sample waveform can ensure better display effect of Lissajous figure.
  • Page 83: Normal

    Chapter 4 To Set the Sample System RIGOL sample points. MSO8000 supports the following four acquisition modes: Normal, Average, Peak, and High Resolution. By default, the acquisition mode is Normal. Press Acquire → Acquisition on the front panel, and then rotate the multifunction knob to select the desired acquisition mode.
  • Page 84: Peak

    Chapter 4 To Set the Sample System RIGOL Figure 4-2 Waveforms before Averaging Figure 4-3 Waveforms after 128 Times of Averaging Peak In this mode, the oscilloscope acquires the maximum and minimum values of the signal within the sample interval to get the envelope of the signal or the narrow pulse that might be lost.
  • Page 85: Sampling Mode

    Chapter 4 To Set the Sample System RIGOL much smoother waveform on the screen and improves the vertical resolution. This is generally used when the sample rate of the digital converter is greater than the storage rate of the acquisition memory.
  • Page 86 Chapter 4 To Set the Sample System RIGOL Note: The sample rate is displayed both in the status bar at the upper part of the screen and in the CH SampleRate menu. To indirectly modify the sample rate, rotate the Horizontal SCALE knob to adjust the horizontal time base or modify the memory depth.
  • Page 87: La Sample Rate

    Chapter 4 To Set the Sample System RIGOL Pulse disappeared LA Sample Rate LA sampling is the process of sampling on the compared digital signals at a certain time interval. LA sample rate is the reciprocal of the time interval. For example, if the LA sample rate is 500 MSa/s, it indicates that the oscilloscope will make data acquisition on the digital signals at an interval of 2 ns.
  • Page 88 Chapter 4 To Set the Sample System RIGOL Trigger Point Pre-sample Delayed Sample Memory Depth Figure 4-4 Memory Depth The following equation shows the relations among memory depth, sample rate, and horizontal time base scale:   MDepth SRate TScale HDivs ——indicates the memory depth.
  • Page 89: La Memory Depth

    Chapter 4 To Set the Sample System RIGOL LA Memory Depth Press Acquire → More to view the memory depth of the current digital channel under LA Mem Depth. The LA memory depth will change with the memory depth of the analog channel, and cannot be set separately.
  • Page 90 Chapter 4 To Set the Sample System RIGOL After you select "User", press Expand User and then rotate the multifunction knob to set the expansion reference. Its available range is from the rightmost position of the screen to the leftmost position of the screen. By default, it is the screen center.
  • Page 91: Chapter 5 To Trigger The Oscilloscope

    Chapter 5 To Trigger the Oscilloscope RIGOL Chapter 5 To Trigger the Oscilloscope As for trigger, you set certain trigger condition according to the requirement and when a waveform in the waveform stream meets this condition, the oscilloscope captures this waveform as well as the neighboring part, and displays them on the screen.
  • Page 92: Trigger Source

    Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Source In the trigger control area (Trigger) on the front panel, press Menu → Source to select the desired source, or rotate the multi-function knob to select a desired source, then press down the knob to select it. You can also press Source continuously to select the trigger source or enable the touch screen to tap the desired trigger source and select it.
  • Page 93: Trigger Level/Threshold Level

    Chapter 5 To Trigger the Oscilloscope RIGOL Trigger LEVEL/Threshold Level The adjustment of the trigger level/threshold level is related to the type of the trigger source. When the trigger source is selected from CH1 to CH4, rotate the Trigger ⚫...
  • Page 94: Trigger Mode

    Chapter 5 To Trigger the Oscilloscope RIGOL Method 1: Enable the touch screen and drag the trigger level line to adjust the trigger level. For details about the operations, refer to descriptions in "Drag" section. Method 2: Enable the touch screen function, and tap the trigger setting label at the upper-right corner of the screen.
  • Page 95 Chapter 5 To Trigger the Oscilloscope RIGOL MSO8000 provides three trigger modes: Auto, Normal, and Single. The default is Auto. In the trigger control area (Trigger) on the front panel, press Mode to quickly switch the current trigger mode. The trigger mode is displayed at the upper-right corner of the screen: A (Auto), N (Normal), and S (Single).
  • Page 96: Trigger Coupling

    Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Coupling Trigger coupling decides which kind of components will be transmitted to the trigger module. Please distinguish it from "Channel Coupling". Note: Trigger coupling is only valid when the trigger type is Edge and the trigger source is an analog channel.
  • Page 97: Noise Rejection

    Chapter 5 To Trigger the Oscilloscope RIGOL In the trigger control area (Trigger) on the front panel, press Menu → Holdoff (for some trigger, Holdoff is a sub menu under More, at this time, press Menu → More → Holdoff) and then rotate the multifunction knob at a small step to modify the holdoff time (the holdoff to this time when the waveforms are stably triggered;...
  • Page 98: Trigger Type

    Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Type MSO8000 series oscilloscope provides the following trigger functions (all the triggers are hardware triggers, except the zone trigger). Edge Trigger ⚫ Pulse Trigger ⚫ Slope Trigger ⚫ Video Trigger ⚫ Pattern Trigger ⚫...
  • Page 99: Edge Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL Edge Trigger Triggers on the trigger level/threshold level of the specified edge of the input signal. Trigger Type: Press Menu → Type in the trigger control area (Trigger) on the front panel and rotate the multifunction knob to select "Edge", then press down the knob to...
  • Page 100: Pulse Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL Trigger LEVEL/Threshold Level: Rotate the Trigger LEVEL knob to adjust the trigger level or threshold level. Please refer to descriptions in "Trigger LEVEL/Threshold Level". The current trigger level/threshold level value is displayed at the upper-right corner of the screen.
  • Page 101 Chapter 5 To Trigger the Oscilloscope RIGOL trigger source, can we obtain a stable trigger. Polarity: Press Polarity continuously to select the desired polarity. The polarities available are positive polarity ( ) and negative polarity ( ). Trigger Condition: Press When, then rotate the multifunction knob to select the trigger condition, and then press down the knob to select the trigger condition.
  • Page 102: Slope Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL the current trigger mode. For details, refer to descriptions in "Trigger Mode". Trigger Parameter Setting: Set the trigger parameters (trigger holdoff and noise rejection) under this trigger type. For details, refer to descriptions in "Trigger Holdoff" and "Noise Rejection".
  • Page 103 Chapter 5 To Trigger the Oscilloscope RIGOL upper-right corner of the screen. Note: Only when we select the channel (that has been input with signals) as the trigger source, can we obtain a stable trigger. Edge Type: Press Slope continuously to select the input signal edge on which the oscilloscope triggers.
  • Page 104 Chapter 5 To Trigger the Oscilloscope RIGOL limit. Level Selection and Adjustment of Trigger Level: After the trigger condition setting is completed, you need to adjust the trigger level to correctly trigger the signal and obtain a stable waveform. Press Level Select (when "<>"...
  • Page 105: Video Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL Video Trigger The video signal can include image information and timing information, which adopts different standards and formats. MSO8000 series can trigger on the standard video signal field or line of NTSC (National Television Standards Committee), PAL (Phase Alternating Line), or SECAM (Sequential Couleur A Memoire).
  • Page 106 Chapter 5 To Trigger the Oscilloscope RIGOL 1080p/50Hz Progressive Scan 1125 1080p/30Hz Progressive Scan 1125 1080p/25Hz Progressive Scan 1125 1080p/24Hz Progressive Scan 1125 1080i/60Hz Interlaced Scan 1125 1080i/50Hz Interlaced Scan 1125 Sync: Press Sync, then rotate the multifunction knob to select the sync type, and then press down the knob to select the type.
  • Page 107: Pattern Trigger

    In the trigger debugging process of video signals, the frequency in different ⚫ part of the signal can be reflected by a different brightness, as RIGOL digital oscilloscope provides the intensity graded color display function. Experienced users can quickly judge the signal quality and discover abnormalities during the debugging process.
  • Page 108 Chapter 5 To Trigger the Oscilloscope RIGOL Pattern Setting: Press Code, and rotate the multifunction knob to select the pattern of the currently selected channel. Press down the knob to select the pattern. You can also press Code continuously to select the pattern or enable the touch screen to select it.
  • Page 109: Duration Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL Note: Only one edge (rising or falling edge) can be specified in the pattern. If one edge item is currently defined and then another edge item is defined in another channel in the pattern, then a prompt message "Invalid input" is displayed.
  • Page 110 Chapter 5 To Trigger the Oscilloscope RIGOL Figure 5-7 Duration Trigger Trigger Type: Press Type, and then rotate the multifunction knob to select "Duration". Press down the knob to select the trigger type. Then, the current trigger setting information is displayed at the upper-right corner of the screen, as shown in the figure below.
  • Page 111 Chapter 5 To Trigger the Oscilloscope RIGOL than the trigger level/threshold level of the channel. L: sets the pattern of the channel selected to "L", i.g. the voltage level is lower ⚫ than the trigger level/threshold level of the channel.
  • Page 112: Timeout Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL Trigger LEVEL/Threshold Level: Rotate the Trigger LEVEL knob to adjust the trigger level/threshold level. Refer to "Trigger LEVEL/Threshold Level". The current trigger level/threshold level value is displayed at the upper-right corner of the screen.
  • Page 113: Runt Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL Rising: starts timing when the rising edge of the input signal passes through ⚫ the trigger level. Falling: starts timing when the falling edge of the input signal passes ⚫ through the trigger level.
  • Page 114 Chapter 5 To Trigger the Oscilloscope RIGOL the knob to select the trigger type. Then, the current trigger setting information is displayed at the upper-right corner of the screen, as shown in the figure below. You can also press Type continuously to select the trigger type or enable the touch screen to tap the desired trigger type and select it.
  • Page 115: Window Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL trigger level remains unchanged. Level AB: adjusts the upper and lower limits of the trigger level at the same time, and the trigger level deviation (the difference between the upper limit and lower limit of the trigger level) remains unchanged.
  • Page 116 Chapter 5 To Trigger the Oscilloscope RIGOL Source Selection: Press Source to switch the trigger source and select CH1-CH4. For details, refer to descriptions in "Trigger Source". The current trigger source is displayed at the upper-right corner of the screen.
  • Page 117: Delay Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL Level AB: adjusts the upper and lower limits of the trigger level at the same time, ⚫ and the trigger level deviation (the difference between the upper limit and lower limit of the trigger level) remains unchanged.
  • Page 118 Chapter 5 To Trigger the Oscilloscope RIGOL Edge A= Edge B= Source A Source B Figure 5-10 Delay Trigger Trigger Type: Press Type, and then use the multifunction knob to select "Delay". Press down the knob to select the trigger type. Then, the current trigger setting information is displayed at the upper-right corner of the screen, as shown in the figure below.
  • Page 119: Setup/Hold Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL >: triggers when the time difference (△T) between the specified edges of ⚫ Source A and Source B is greater than the preset time lower limit. Press Lower to set the delay time lower limit in Delay trigger.
  • Page 120 Chapter 5 To Trigger the Oscilloscope RIGOL T1 is the setup time is the hold time Data Source Clock Source The data type is H The edge type is Figure 5-11 Setup/Hold Trigger Trigger Type: Press Type, and then rotate the multifunction knob to select "Setup/Hold".
  • Page 121 Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Condition: Press When and rotate the multifunction knob to select the trigger condition of Setup/Hold trigger. Press down the knob to select the trigger condition. You can also press When continuously or enable the touch screen to select it.
  • Page 122: Nth Edge Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL Nth Edge Trigger Triggers on the Nth edge that appears after the specified idle time. For example, in the waveform shown in the figure below, the instrument should trigger on the second rising edge after the specified idle time (the time between two neighboring rising edges), and the idle time should be within the range between P and M (P<Idle...
  • Page 123: Rs232 Trigger (Option)

    Chapter 5 To Trigger the Oscilloscope RIGOL Idle Time: Press Idle Time to set the idle time before the edge counting in Nth edge trigger. Rotate the multifunction knob or use the pop-up numeric keypad to set the idle time.
  • Page 124 Chapter 5 To Trigger the Oscilloscope RIGOL enable the touch screen to tap the desired trigger type and select it. Source: Press Source to open the signal source list and select CH1-CH4 or D0-D15. For details, refer to descriptions in "Trigger Source". The current trigger source is displayed at the upper-right corner of the screen.
  • Page 125 Chapter 5 To Trigger the Oscilloscope RIGOL Baud Rate: Set the baud rate of data transmission (i.g. specifies a clock frequency). Three setting methods are available for baud rate: Press Baud Rate, then set the user-defined baud rate with the displayed ⚫...
  • Page 126: I2C Trigger (Option)

    Chapter 5 To Trigger the Oscilloscope RIGOL I2C Trigger (Option) I2C is a 2-wire serial bus used to connect the microcontroller and its peripheral device. It is a bus standard widely used in the microelectronic communication control field. The I2C serial bus consists of SCL and SDA. Its transmission rate is determined by SCL, and its transmission data is determined by SDA, as shown in the figure below.
  • Page 127 Chapter 5 To Trigger the Oscilloscope RIGOL Start: triggers when SDA data transitions from high level to low level while SCL is ⚫ high level. Stop: triggers when SDA data transitions from low level to high level while SCL is ⚫...
  • Page 128 Chapter 5 To Trigger the Oscilloscope RIGOL Press Bytes, then rotate the multifunction knob or use the pop-up ➢ numeric keypad to set the length of the data. Its range is from 1 to 5. Press More → Addr Bits, and rotate the multifunction knob to select ➢...
  • Page 129: Spi Trigger (Option)

    Chapter 5 To Trigger the Oscilloscope RIGOL SPI Trigger (Option) In SPI trigger, after the CS or timeout condition is satisfied, the oscilloscope triggers when the specified data is found. When using SPI trigger, you need to specify the SCL clock sources (CLK) and SDA data sources (MISO). Below is the sequential chart of SPI bus.
  • Page 130 Chapter 5 To Trigger the Oscilloscope RIGOL time. After selecting this condition, you can press Timeout, then rotate the multifunction knob or use the pop-up numeric keypad to set the idle time. The range is from 8 ns to 10 s.
  • Page 131: Can Trigger (Option)

    Chapter 5 To Trigger the Oscilloscope RIGOL CAN Trigger (Option) MSO8000 can trigger on the start of a frame, end of a frame, frame of the specified type (e.g. Remote, Overload, Data, etc.), or error frame of the specified type (e.g.
  • Page 132 Chapter 5 To Trigger the Oscilloscope RIGOL Frame ID: triggers on the data frames with the specified ID. Set the following ⚫ parameters: Press Extended ID to enable or disable the extended ID. ➢ Press Bit X to set the data bit that needs to be operated on. For setting ➢...
  • Page 133 Chapter 5 To Trigger the Oscilloscope RIGOL Sample Position: Sample position is a point within a bit’s time. The oscilloscope samples the bit level at this point. The sample position is represented by the proportion of "the time from the start of the bit to the sample position"...
  • Page 134: Flexray Trigger (Option)

    Chapter 5 To Trigger the Oscilloscope RIGOL FlexRay Trigger (Option) MSO8000 can trigger on the specified frame, symbol, error, or position of the FlexRay bus. FlexRay is a type of differential serial bus configured with three consecutive segments (i.g. header, payload, and trailer). Its data transmission rate is up to 10 Mb/s.
  • Page 135 Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Condition: Press When, and rotate the multifunction knob to select the desired trigger condition, then press down the knob to select it. You can also press When continuously or enable the touch screen to select it.
  • Page 136: Lin Trigger (Option)

    Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Parameter Setting: Set the trigger parameter (noise rejection) under this trigger type. For details, refer to descriptions in "Noise Rejection". Trigger LEVEL/Threshold Level: Rotate the Trigger LEVEL knob to adjust the trigger level/threshold level. Refer to "Trigger LEVEL/Threshold Level".
  • Page 137 Chapter 5 To Trigger the Oscilloscope RIGOL keypad to set ID. Data: triggers when the data that meet the preset conditions are found. ⚫ Press Bit X to set the data bit that needs to be operated on. For setting ➢...
  • Page 138 Chapter 5 To Trigger the Oscilloscope RIGOL One Bit Figure 5-21 Sample Position Press More → Sample Position, then rotate the multifunction knob or use the pop-up numeric keypad to set the sample position. The settable range is from 10% to 90%.
  • Page 139: I2S Trigger (Option)

    Chapter 5 To Trigger the Oscilloscope RIGOL I2S Trigger (Option) In I2S trigger, the oscilloscope searches for the specified data value and take it as the condition for identifying the trigger. You need to specify the serial clock line (SCLK, 1...
  • Page 140 Chapter 5 To Trigger the Oscilloscope RIGOL Audio: Press Audio continuously or rotate the multifunction knob to select "Left", "Right", or "Either". Trigger Condition: Press More → When, and rotate the multifunction knob to select the desired trigger condition, then press down the knob to select it. You can also press When continuously or enable the touch screen to select it.
  • Page 141 Chapter 5 To Trigger the Oscilloscope RIGOL Significant Bit) is sent last. The MSB appears on the SDA line one bit clock after the edge of the WS transition. LJ: data transmission (MSB first) begins at the edge of the WS transition.
  • Page 142: Mil-Std-1553 Trigger (Option)

    Chapter 5 To Trigger the Oscilloscope RIGOL MIL-STD-1553 Trigger (Option) 1553B is the abbreviation for the MIL-STD-1553 bus. MSO8000 can trigger on the sync field of 1553B bus, and can also trigger on the specified data word, command word, status word, or error type.
  • Page 143 Chapter 5 To Trigger the Oscilloscope RIGOL descriptions in "Trigger Source". The current trigger source is displayed at the upper-right corner of the screen. Note: Only when we select the channel (that has been input with signals) as the trigger source, can we obtain a stable trigger.
  • Page 144 Chapter 5 To Trigger the Oscilloscope RIGOL press Err Type to select the error type. Sync Error: triggers when an invalid sync pulse is found. ➢ Check Error: triggers when the parity bit is incorrect for the data in the ➢...
  • Page 145: Zone Trigger

    Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Parameter Setting: Set the trigger parameter (noise rejection) under this trigger type. For details, refer to descriptions in "Noise Rejection". Zone Trigger MSO8000 series oscilloscope supports the zone trigger and provides two rectangle areas: Trigger zone A and Trigger zone B.
  • Page 146 Chapter 5 To Trigger the Oscilloscope RIGOL If you enable Zone A, select CH1 as Source A, and select "Not intersect" as the condition, the following figure is displayed. If you enable both Trigger zone A and Trigger zone B, then perform the "A&&B"...
  • Page 147: Trigger Output Connector

    Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Output Connector The trigger output connector ([TRIG OUT]) on the rear panel of MSO8000 series can output trigger signals (hardware trigger) determined by the current setting. Trigger Output Connector Figure 5-24 Trigger Output Connector Press Utility →...
  • Page 149: Chapter 6 Operations And Measurements

    Chapter 6 Operations and Measurements RIGOL Chapter 6 Operations and Measurements MSO8000 can perform math operations, auto measurements, and cursor measurements on sampled and displayed data. Contents in this chapter: Math Operation ◼ Auto Measurement ◼ Cursor Measurement ◼ MSO8000 User Guide...
  • Page 150: Math Operation

    Chapter 6 Operations and Measurements RIGOL Math Operation Press Math in the waveform control area (Wave) on the front panel to open the math operation menu. You can also tap the function navigation icon at the lower-left corner of the screen to open the function navigation. Then tap the "Math"...
  • Page 151: Subtraction

    Chapter 6 Operations and Measurements RIGOL to select the channel. You can also press SourceA and SourceB continuously or enable the touch screen to select the channel. The available channels for SourceA and SourceB are CH1-CH4 and Ref1-Ref10. Note: The source of Math2 can be CH1-CH4, Ref1-Ref10, or Math1; the source of Math3 can be CH1-CH4, Ref1-Ref10, Math1, or Math2;...
  • Page 152: Multiplication

    Chapter 6 Operations and Measurements RIGOL SourceA and SourceB are CH1-CH4 and Ref1-Ref10. Note: The source of Math2 can be CH1-CH4, Ref1-Ref10, or Math1; the source of Math3 can be CH1-CH4, Ref1-Ref10, Math1, or Math2; the source of Math4 can be CH1-CH4, Ref1-Ref10, Math1, Math2, or Math3; only the channels that have been enabled currently can be selected.
  • Page 153: Division

    Chapter 6 Operations and Measurements RIGOL of Math3 can be CH1-CH4, Ref1-Ref10, Math1, or Math2; the source of Math4 can be CH1-CH4, Ref1-Ref10, Math1, Math2, or Math3; only the channels that have been enabled currently can be selected. Press Offset to set the vertical offset of the operation results. For setting ⚫...
  • Page 154: Fft

    Chapter 6 Operations and Measurements RIGOL of Math3 can be CH1-CH4, Ref1-Ref10, Math1, or Math2; the source of Math4 can be CH1-CH4, Ref1-Ref10, Math1, Math2, or Math3; only the channels that have been enabled currently can be selected. Press Offset to set the vertical offset of the operation results. For setting ⚫...
  • Page 155 Chapter 6 Operations and Measurements RIGOL the following figure. Of which, FFT resolution is the quotient of the sample rate and the number of FFT points. If the number of FFT points is a fixed value (65535 at most), then the lower the sample rate, the higher the resolution.
  • Page 156 Chapter 6 Operations and Measurements RIGOL the frequency at the right side of the screen. Press End to set the stop frequency of the frequency-domain waveform. For setting methods, refer to descriptions in "Parameter Setting Method". Press More → Peak Search to enter the peak search menu.
  • Page 157 Chapter 6 Operations and Measurements RIGOL Table 6-1 Window Function Window Waveforms Applicable to the Characteristics Function Window Function Rectangular Best frequency Transient or short pulse, the signal resolution levels before and after the Poorest multiplication are basically the same.
  • Page 158: And" Operation

    Chapter 6 Operations and Measurements RIGOL FFT shortcut operation: Enable the touch screen, and then tap the function navigation icon at the lower-left corner of the screen to open the function navigation. Then, tap the "FFT" icon to open the FFT menu. This oscilloscope supports 4 FFTs: FFT1, FFT2, FFT3, and FFT4.
  • Page 159: Or" Operation

    Chapter 6 Operations and Measurements RIGOL threshold of Source A (or Source B) in logic operation. For setting methods, refer to descriptions in "Parameter Setting Method". The thresholds of the digital channels (D0-D15) are respectively set in the LA module.
  • Page 160: Xor" Operation

    Chapter 6 Operations and Measurements RIGOL Press SourceA and SourceB respectively, and rotate the multifunction knob ⚫ to select the channel for SourceA and Source B. Then press down the knob to select the channel. You can also press SourceA and SourceB continuously or enable the touch screen to select the channel.
  • Page 161 Chapter 6 Operations and Measurements RIGOL two binary bits are as shown in Table 6-4. Table 6-4 Logic "XOR" Operation Press Math → Math1 → Operator to select "A^B": Press Operation continuously to enable or disable the "XOR" operation ⚫...
  • Page 162: Not" Operation

    Chapter 6 Operations and Measurements RIGOL from the analog signal on the source. For setting methods, refer to descriptions in "Parameter Setting Method". Press More → Label to set the label for math operation results. For details, ⚫ refer to descriptions in "Math Operation Label".
  • Page 163: Intg

    Chapter 6 Operations and Measurements RIGOL Press Size, and then rotate the multifunction knob to select the waveform ⚫ size. Press down the knob to select it. You can also press Size continuously or enable the touch screen to select it. The available sizes include "Small", "Medium", and "Large".
  • Page 164: Diff

    Chapter 6 Operations and Measurements RIGOL Press More → Invert to enable or disable the inverted display function of the ⚫ waveform. Press More → Label to set the label for math operation results. For details, ⚫ refer to descriptions in "Math Operation Label".
  • Page 165: Sqrt

    Chapter 6 Operations and Measurements RIGOL Press More → Invert to enable or disable the inverted display function of the ⚫ waveform. Press More → Label to set the label for math operation results. For details, ⚫ refer to descriptions in "Math Operation Label".
  • Page 166: Lg (Base 10 Exponential)

    Chapter 6 Operations and Measurements RIGOL waveform. Press More → Label to set the label for math operation results. For details, ⚫ refer to descriptions in "Math Operation Label". Press More → Color Grade to enable or disable the color grade display of the ⚫...
  • Page 167 Chapter 6 Operations and Measurements RIGOL Press More → Reset ColorGrade to clear the color grade display and display ⚫ the color grade again. Calculates the natural logarithm (Ln) of the selected source point by point and displays the results.
  • Page 168: Exp

    Chapter 6 Operations and Measurements RIGOL Calculates the exponential of the selected source point by point and displays the results. Press Math → Math1 → Operator to select "Exp": Press Operation continuously to enable or disable the "Exp" operation function.
  • Page 169: Low Pass

    Chapter 6 Operations and Measurements RIGOL Press Operation continuously to enable or disable the "Abs" operation function. ⚫ Press SourceA, and rotate the multifunction knob to select the channel for ⚫ SourceA. Then press down the knob to select the channel. You can also press SourceA continuously or enable the touch screen to select the channel.
  • Page 170: High Pass

    Chapter 6 Operations and Measurements RIGOL available channels for SourceA are CH1-CH4 and Ref1-Ref10. Note: The source of Math2 can be CH1-CH4, Ref1-Ref10, or Math1; the source of Math3 can be CH1-CH4, Ref1-Ref10, Math1, or Math2; the source of Math4 can be CH1-CH4, Ref1-Ref10, Math1, Math2, or Math3; only the channels that have been enabled currently can be selected.
  • Page 171: Band Pass

    Chapter 6 Operations and Measurements RIGOL SourceA continuously or enable the touch screen to select the channel. The available channels for SourceA are CH1-CH4 and Ref1-Ref10. Note: The source of Math2 can be CH1-CH4, Ref1-Ref10, or Math1; the source of Math3 can be CH1-CH4, Ref1-Ref10, Math1, or Math2; the source of Math4 can be CH1-CH4, Ref1-Ref10, Math1, Math2, or Math3;...
  • Page 172: Band Stop

    Chapter 6 Operations and Measurements RIGOL SourceA. Then press down the knob to select the channel. You can also press SourceA continuously or enable the touch screen to select the channel. The available channels for SourceA are CH1-CH4 and Ref1-Ref10.
  • Page 173 Chapter 6 Operations and Measurements RIGOL Press Operation continuously to enable or disable the "BandStop" operation ⚫ function. Press SourceA, and rotate the multifunction knob to select the channel for ⚫ SourceA. Then press down the knob to select the channel. You can also press SourceA continuously or enable the touch screen to select the channel.
  • Page 174: Ax+B

    Chapter 6 Operations and Measurements RIGOL AX+B Applies a linear function to the selected source, and displays the results. Press Math → Math1 → Operator to select "AX+B": Press Operation continuously to enable or disable the "AX+B" operation ⚫ function.
  • Page 175: Avg

    Chapter 6 Operations and Measurements RIGOL Averages the waveforms of the specified channel for several times and displays the results. Press Math → Math1 → Operator to select "Avg": Press Operation continuously to enable or disable the "Avg" operation function.
  • Page 176: Math Operation Label

    Chapter 6 Operations and Measurements RIGOL indicate the times of data acquisition or acquisition probability. Press More → Reset ColorGrade to clear the color grade display and display ⚫ the color grade again. Math Operation Label Press Label to enter the label setting menu.
  • Page 177: Auto Measurement

    Chapter 6 Operations and Measurements RIGOL Auto Measurement MSO8000 provides auto measurements for 41 waveform parameters, as well as the statistics and analysis of the measurement results. Quick Measurement after AUTO When the oscilloscope is correctly connected and has detected a valid input signal, press AUTO to enable the waveform auto setting function and open the auto setting function menu.
  • Page 178 Chapter 6 Operations and Measurements RIGOL If you lock the Auto function and then exit the auto setting function menu, you can press Utility → More → Auto Config to enter the auto setting function menu again to unlock the Auto function.
  • Page 179 Chapter 6 Operations and Measurements RIGOL to 5 mV. If not meeting the conditions, the waveform auto setting function may be invalid. MSO8000 User Guide 6-31...
  • Page 180: Measurement Parameter

    Chapter 6 Operations and Measurements RIGOL Measurement Parameter Press Measure on the front panel to enter the measurement setting menu. You can set the measurement source, enable or disable the all measurement function, the statistical analysis function, and etc. You can also quickly make measurements for 41 waveform parameters.
  • Page 181 Chapter 6 Operations and Measurements RIGOL 6. -Width: indicates the time between the threshold middle value of a falling edge to the threshold middle value of the next rising edge. 7. +Duty: indicates the ratio of the positive pulse width to the period.
  • Page 182 Chapter 6 Operations and Measurements RIGOL Rising Edge Count = n Threshold Upper Limit Threshold Lower Limit 4. Falling Edge Count: The number of falling edges that fall from above the threshold upper limit to below the threshold lower limit.
  • Page 183 Chapter 6 Operations and Measurements RIGOL indicates that the falling edge of Source A occurred after that of Source B. 3. Delay A →B : indicates the time difference between the threshold middle values of the rising edge of Source A and the falling edge of Source B. Negative delay indicates that the rising edge of Source A occurred after the falling edge of Source B.
  • Page 184 Chapter 6 Operations and Measurements RIGOL Note: Source A and Source B can be any channel among CH1-CH4, Math1-Math4, or ⚫ D0-D15. You can press Measure → Add to enter the sub-menu to set the parameters. The default threshold middle value is 50%. You can press Measure → Setting ⚫...
  • Page 185 Chapter 6 Operations and Measurements RIGOL gating area. The formula is shown as follows: ∑ Average Wherein, is the measurement result of the i th point, and n is the number of points being measured. 11. VRMS: indicates the root mean square value on the whole waveform or in the gating area.
  • Page 186: Measurement Settings

    Chapter 6 Operations and Measurements RIGOL The unit is V*s. The area of the waveform above the zero reference (namely the vertical offset) is positive, and the area of the waveform below the zero reference is negative. The area measured is the algebraic sum of the whole period area.
  • Page 187 Chapter 6 Operations and Measurements RIGOL rightmost item at the bottom of the screen will be disabled. You can also enable the touch screen to select the measurement category and the measurement parameters. When you rotate the multifunction knob to switch from the ⚫...
  • Page 188 Chapter 6 Operations and Measurements RIGOL resolution of measurement results. Note, in this mode, the refresh rate of the waveforms may be reduced. Press Type to select "Threshold", "Range", or "Amp Method". ⚫ Select "Threshold", and then make the following settings: ➢...
  • Page 189: Remove The Measurement Result

    Chapter 6 Operations and Measurements RIGOL set the following parameters: --Press Top to select "Histogram" or "Max-Min". --Press Base to select "Histogram" or "Max-Min". Modifying the threshold will affect the measurement results of ⚫ the time, delay and phase parameters.
  • Page 190: All Measurement

    Chapter 6 Operations and Measurements RIGOL Press Measure → Statistic to enter the sub-menu of "Statistic". Press Statistic to enable or disable the statistical function. ⚫ Press Reset Stat. to clear the history statistics data and makes statistics again. ⚫...
  • Page 191 Chapter 6 Operations and Measurements RIGOL X Cursor (Cursor A) X Cursor (Cursor B) Y Cursor (Cursor A) Y Cursor (Cursor B) X Cursor ⚫ X cursor is a vertical solid/dotted line that is used to make horizontal adjustments. It can be used to measure time (s), frequency (Hz), phase (°), and ratio (%).
  • Page 192: Manual Mode

    Chapter 6 Operations and Measurements RIGOL Note: When the Time Base Mode is set to "XY", the XY cursor measurement mode is ➢ valid. If no measurement parameter is selected or no measurement source is input, no ➢ cursor is displayed.
  • Page 193 Chapter 6 Operations and Measurements RIGOL Y: It is a pair of horizontal solid (Cursor A)/dotted (Cursor B) lines, used for ⚫ measuring voltage parameters. The measurement results include AY, BY, and △Y. 2. Select Measurement Source Press Source, then rotate the multifunction knob to select the desired source (CH1-CH4).
  • Page 194 Chapter 6 Operations and Measurements RIGOL Press BY, then rotate the multifunction knob to adjust the vertical ⚫ position of Cursor B (Y cursor). Enable the touch screen and tap BY, then the numeric keypad is displayed. Set the vertical position of Cursor B (Y cursor). The vertical axis indicates voltage, and the unit of its setting value is the same as that of the vertical axis.
  • Page 195 Chapter 6 Operations and Measurements RIGOL Degree(°): measures the phase value at X Cursor. The measurement ➢ results include AX, BX, and △X, and they are all expressed in degree. This oscilloscope allows you to set the reference position for the phase according to your needs.
  • Page 196: Track Mode

    Chapter 6 Operations and Measurements RIGOL When "LA" is selected under Source, you need to set the LA display type. Press More and then press La Type continuously to select "HEX" or "BIN". 7. Measurement Example Measure the period of a sine wave by using the manual cursor measurement and auto measurement respectively.
  • Page 197 Chapter 6 Operations and Measurements RIGOL AX: indicates the X value at Cursor A. ⚫ AY: indicates the Y value at Cursor A. ⚫ BX: indicates the X value at Cursor B. ⚫ BY: indicates the Y value at Cursor B.
  • Page 198 Chapter 6 Operations and Measurements RIGOL between Cursor A and Cursor B (X cursors) remains unchanged. 2) When "Y" is selected under Track, you can adjust the position of Y cursor. Press AY, then rotate the multifunction knob to adjust the vertical ⚫...
  • Page 199: Xy Mode

    Chapter 6 Operations and Measurements RIGOL XY Mode XY mode is only available when the horizontal time base mode is "XY" (refer to "Timebase Mode"). In this mode, two pairs of cursors will appear. You can adjust the cursor positions to measure the X and Y values at the crossing points of the two pairs of cursors.
  • Page 200 Chapter 6 Operations and Measurements RIGOL Cursor B. Press More → ABX and rotate the multifunction knob to adjust the X ⚫ value at Cursor A and the X value at Cursor B simultaneously. Press More → ABY and rotate the multifunction knob to adjust the Y ⚫...
  • Page 201: Measure Mode

    Chapter 6 Operations and Measurements RIGOL Measure Mode Press Mode to select "Measure". In this mode, the cursor position for measurement is displayed. Press Indicator continuously to enable or disable the indicator. If it is enabled, one or multiple cursors are displayed on the screen. Before enabling the indicator, you need to enable at least one parameter for auto measurement.
  • Page 203: Chapter 7 Digital Voltmeter (Dvm) And Frequency Counter

    Chapter 7 Digital Voltmeter (DVM) and Frequency Counter RIGOL Chapter 7 Digital Voltmeter (DVM) and Frequency Counter MSO8000 series oscilloscope provides a built-in digital voltmeter (DVM) and frequency counter, which enable you to perform an accurate measurement, improving user experience in counter and frequency measurement.
  • Page 204: Digital Voltmeter (Dvm)

    Chapter 7 Digital Voltmeter (DVM) and Frequency Counter RIGOL Digital Voltmeter (DVM) The built-in DVM of this oscilloscope provides 3-digit voltage measurements on any analog channel. DVM measurements are asynchronous from the oscilloscope’s acquisition system and are always acquiring. Press Measure → Analyze → DVM to open the DVM setting menu, or enable the...
  • Page 205: To Select Measurement Mode

    Chapter 7 Digital Voltmeter (DVM) and Frequency Counter RIGOL (CH1-CH4) can be selected to be the measurement source. Note: Even if the analog channel (CH1-CH4) is not enabled, you can still perform the DVM measurement. To Select Measurement Mode Press Mode in the "DVM" setting menu, then rotate the multifunction knob select the desired mode, and then press down the knob to select the mode.
  • Page 206: Frequency Counter

    Chapter 7 Digital Voltmeter (DVM) and Frequency Counter RIGOL Frequency Counter The frequency counter analysis function provides frequency, period, or edge event counter measurements on any analog channel. Press Measure → Analyze → Counter to open the frequency counter setting menu, or enable the touch screen and then tap the function navigation icon the lower-left corner of the screen to open the function navigation.
  • Page 207: To Set Resolution

    Chapter 7 Digital Voltmeter (DVM) and Frequency Counter RIGOL Press Measure in the "Counter" menu, then rotate the multifunction knob select the desired measurement item, and then press down the knob to select the item. You can also press Measure continuously or enable the touch screen to tap the desired measurement item and select it.
  • Page 209: Chapter 8 Power Analysis (Option)

    Chapter 8 Power Analysis (Option) RIGOL Chapter 8 Power Analysis (Option) MSO8000 series oscilloscope provides an optional power analysis function, which can help you easily analyze the efficiency and reliability of the switch-mode power supplies. With the power analysis function, you can analyze the power quality and output ripple noise of the input power.
  • Page 210: Power Quality

    Chapter 8 Power Analysis (Option) RIGOL Power Quality By analyzing the power quality, you can test the quality of AC input lines. The specific measurement parameters for power quality analysis include V_RMS, I_RMS, real power, apparent power, reactive power, power factor, reference frequency, phase angle, impedance, voltage crest factor, and current crest factor.
  • Page 211 Chapter 8 Power Analysis (Option) RIGOL numeric keypad to set the upper limit value. Press Mid, then rotate the multifunction knob or use the pop-up ➢ numeric keypad to set the middle value. The middle value is limited by the settings of the upper limit and lower limit.
  • Page 212: Ripple

    Chapter 8 Power Analysis (Option) RIGOL Figure 8-1 Connection Diagram of Power Quality Analysis Ripple Power ripple is an important parameter for evaluating DC power supply, which indicates the ripple quantity of the output DC voltage. The ripple analysis can measure the current value, average value, minimum value, maximum value, standard deviation, and count value of the ripples on the power output terminal.
  • Page 213 Chapter 8 Power Analysis (Option) RIGOL the figure below. You can also enable the touch screen to tap the icon at the right-upper corner of the statistical results window to close it. View the Connection Diagram ⚫ Press Tips, and then the connection diagram of the ripple analysis is displayed on the screen.
  • Page 215: Chapter 9 Histogram Analysis

    Chapter 9 Histogram Analysis RIGOL Chapter 9 Histogram Analysis MSO8000 series oscilloscope supports the histogram analysis function, enabling you to judge the trend of waveforms, and quickly locate the potential problems of the signal. Contents in this chapter: To Enable or Disable the Histogram Function ◼...
  • Page 216: To Enable Or Disable The Histogram Function

    Chapter 9 Histogram Analysis RIGOL To Enable or Disable the Histogram Function The histogram analysis function supported by the MSO8000 series is available to provide a statistical view of waveforms or measurement results. It can be classified into horizontal histogram, vertical histogram, measurement histogram, and jitter histogram.
  • Page 217: To Select The Histogram Source

    Chapter 9 Histogram Analysis RIGOL graticule. Press Type, then rotate the multifunction knob to select the desired histogram type, and then press down the knob to select the type. You can also press Type continuously or enable the touch screen to select the type.
  • Page 218: To Enable Or Disable The Statistical Function

    Chapter 9 Histogram Analysis RIGOL To Enable or Disable the Statistical Function Press Statistic to enable or disable the statistical function. When the statistical function is enabled, the statistical results of the histogram data will be displayed, as shown in the following figure. Note that when you select "Measure" under Type, the statistical data results also include the parameter XScale.
  • Page 219: Chapter 10 Real-Time Eye Diagram And Jitter Analysis (Option)

    Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) The MSO8000 series oscilloscope provides the real-time eye plot and measurement with the clock recovery function. The flexible and convenient jitter measurement and...
  • Page 220: Real-Time Eye Analysis

    Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL Real-time Eye Analysis An eye diagram is a view of a signal. A real- time eye accomplishes this by acquiring data, performing clock recovery, and then superimposing (folding) successive unit intervals within a single plot.
  • Page 221: To Set The Threshold

    Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL To Set the Threshold In the "Eye" setting menu, press Threshold to enter the threshold setting menu. Set the high threshold ⚫ Press HighThres, then rotate the multifunction knob or use the pop-up numeric keypad to set the high threshold.
  • Page 222 Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL Press Type, then rotate the multifunction knob to select the desired data rate type. Press down the knob to select it. You can also press the Type key continuously or enable the touch screen to select it.
  • Page 223: To Enable Or Disable The Eye Measurement Result

    Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL Press Source, then rotate the multifunction knob to select the desired source and press down the knob to select it. You can also press Source continuously or enable the touch screen to select it. The analog channel (CH1-CH4) can be selected to be the source of the external clock.
  • Page 224: To Reset Color

    Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL Figure 10-1 Diagram of Eye Measurement Parameters To Reset Color Press Reset Color to clear the color level counter. Jitter Analysis The jitter analysis function is mainly used to analyze the integrity of the high-speed serial signal and measure the variance of a measurement over time.
  • Page 225: To Enable Or Disable The Jitter Analysis Function

    Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL To Enable or Disable the Jitter Analysis Function In the "Jitter" setting menu, press the Enable key continuously to enable or disable the jitter analysis function. When the jitter analysis function and the jitter trend are enabled, Math3 trend operation function is enabled by default and the trend diagram is displayed on the screen.
  • Page 226 Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL ➢ Compares edges in a signal under test with the edges in an ideal signal determined by the clock recovery feature. Then, all the signal intervals are measured based on the ideal data rate to generate the error statistics.
  • Page 227 Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL width for the second measurement result, and so on, until all of the cycles of the waveform have been measured. Jitter edge ⚫ When the measurement item is "TIE" or "Cycle to Cycle", you need to set the edge for the measurement signal.
  • Page 228: To Enable Or Disable The Jitter Measurement Result

    Chapter 10 Real-time Eye Diagram and Jitter Analysis (Option) RIGOL Note: When you select the specified measurement item and enable the display of the histogram, the selected item will be displayed under the Item menu of the Histogram setting menu simultaneously. Meanwhile, the jitter analysis function in the histogram setting menu will also be enabled.
  • Page 229: Chapter 11 Digital Channel

    Chapter 11 Digital Channel RIGOL Chapter 11 Digital Channel MSO8000 series oscilloscope has a standard configuration of the logic analyzer (LA) function, and it has 16 digital channels. The default channel label is D15-D0. The oscilloscope compares the voltages acquired in each sample with the preset logic threshold.
  • Page 230: To Select The Digital Channel

    Chapter 11 Digital Channel RIGOL To Select the Digital Channel Press LA on the front panel to enter the LA setting menu. Press Select to open the channel menu, and rotate the multifunction knob to select the channel. You can also press Select continuously or enable the touch screen to select it.
  • Page 231: To Set The Threshold

    Chapter 11 Digital Channel RIGOL D15-D8 channels. Enable/disable a single digital channel ⚫ Press Dx On/Off to open the channel selection list. Rotate the multifunction knob to select a channel, and then press down the knob to enable or disable it.
  • Page 232: Auto Arrangement Setting

    Chapter 11 Digital Channel RIGOL Besides, you can also enable the touch screen to tap the digital channel status area at the bottom of the screen. Then, the LA window is displayed. In the window, you can select the LA channel and set the threshold. Tap the icon at the right side of the threshold field, and then the threshold selection list is displayed.
  • Page 233: Group Setting

    Chapter 11 Digital Channel RIGOL Use preset labels ⚫ Press Library to select a preset label. The available preset labels include ACK, ADO, ADDR, BIT, CAS, CLK, CS, DATA, HALT, INT, LOAD, NMI, OUT, PAS, PIN, RDY, RST, RX, TX, WR, MISO, MOSI, and D0-D15.
  • Page 234: Waveform Color Of The Digital Channel

    Chapter 11 Digital Channel RIGOL Waveform Color of the Digital Channel When the digital channel is enabled, its corresponding waveforms and label are displayed on the screen. This oscilloscope supports setting high logic level display color, low logic level display color, and edge display color.
  • Page 235: Chapter 12 Protocol Decoding

    Chapter 12 Protocol Decoding RIGOL Chapter 12 Protocol Decoding You can use the protocol analysis to discover errors, debug hardware, and accelerate development easily, ensuring you to accomplish the projects with high speed and good quality. Protocol decoding is the basis of protocol analysis. Only protocol analyses with correct protocol decoding are acceptable, and only correct protocol decoding can identify more error information.
  • Page 236: Parallel Decoding

    Chapter 12 Protocol Decoding RIGOL Parallel Decoding Parallel bus consists of clock line and data line. As shown in the figure below, CLK is the clock line, whereas Bit0 and Bit1 are the 0 bit and 1st bit on the data line respectively.
  • Page 237 Chapter 12 Protocol Decoding RIGOL Falling : samples the channel data on the falling edge of the clock. Both : samples the channel data on both the rising edge and the falling edge of the clock. Set the threshold ⚫...
  • Page 238 Chapter 12 Protocol Decoding RIGOL D8-D15 8 0 (default) Bit0 to Bit7 are set to D8 to D15 (default) respectively. Width is set automatically, and you cannot modify them. D0-D15 16 0 (default) Bit0 to Bit15 are set to D0 to (default) D15 respectively.
  • Page 239 Chapter 12 Protocol Decoding RIGOL Polarity ⚫ Press Polarity continuously to select Positive or Negative as the polarity for data decoding. 4. Display-related setting Press Display to enter the display setting menu. Set the display format ⚫ Press Format, rotate the multifunction knob to select the display format of the bus data.
  • Page 240 Chapter 12 Protocol Decoding RIGOL Line Number Time Decode Name Decoded Data Event Table Parallel Bus Figure 12-2 Parallel Decoding Event Table Note: When you adjust the horizontal time base, the waveform displayed ⚫ on the screen will also change, and the total number of lines containing the decoding information in the event table will also be changed.
  • Page 241 Chapter 12 Protocol Decoding RIGOL the decoded data are not fully displayed. At this time, you can view the details in the "Packets" view. Export: If you select "Packets", you can export time and the ➢ corresponding decoded data. Press Export, the save setting menu is displayed.
  • Page 242: Rs232 Decoding (Option)

    Chapter 12 Protocol Decoding RIGOL RS232 Decoding (Option) RS232 serial bus consists of the transmitting data line (TX) and the receiving data line (RX). DeviceA DeviceB Figure 12-3 Schematic Diagram of RS232 Serial Bus The industry standard of RS232 uses "Negative Logic", i.g. high level is logic "0" and low level is logic "1".
  • Page 243 Chapter 12 Protocol Decoding RIGOL RS232 decoding function (set the corresponding RS232 decoding parameters automatically). For source channels, copy the source channel of the trigger to the Tx source channel of the decoding. For analog channels, the threshold level settings will also be copied.
  • Page 244 Chapter 12 Protocol Decoding RIGOL Press Data and rotate the multifunction knob to set the number of data bits actually contained in each frame of data. Press down the knob to select it. You can also press Data continuously or enable the touch screen to select it.
  • Page 245 Chapter 12 Protocol Decoding RIGOL Press Separator and rotate the multifunction knob to select the desired package separator in data transmission, then press down the knob to select it. You can also press Separator continuously or enable the touch screen to select it. It can be set to 0A(LF), 0D(CR), 20(SP), or 0(NUL).
  • Page 246 Chapter 12 Protocol Decoding RIGOL The displayed decoded data information in the bus is related to the ⚫ value of the horizontal time base. Reducing the horizontal time base can help you view the detailed information. Set the event table format: Press Format, rotate the multifunction ➢...
  • Page 247 Chapter 12 Protocol Decoding RIGOL information will be displayed. For example, when "none" is chosen for the parity on the transmitter and "odd" is chosen for the parity on the decoder, the following parity error information (its display form is related to the value of the horizontal time base) occurs.
  • Page 248 Chapter 12 Protocol Decoding RIGOL 12-14 MSO8000 User Guide...
  • Page 249: I2C Decoding (Option)

    Chapter 12 Protocol Decoding RIGOL I2C Decoding (Option) I2C serial bus consists of the clock line (SCLK) and the data line (SDA). SCL: samples SDA on the rising or falling edge of the clock. SDA: indicates the data channel. Figure 12-5 I2C Serial Bus In the decode setting menu, press Decode1 →...
  • Page 250 Chapter 12 Protocol Decoding RIGOL Exchange sources ⚫ Press Exchange to select "SCL/SDA" or "SDA/SCL" to exchange the sources of the current clock channel and data channel. 4. Specify whether the address information includes the "R/W" bit For I2C bus, each frame of data starts with the address information. The address information includes the read address and the write address.
  • Page 251 Chapter 12 Protocol Decoding RIGOL Line Number Time R/W Bit Address Decoded Data Acknowledgement Event Table I2C Bus Figure 12-6 I2C Decoding Event Table Note: When you adjust the horizontal time base, the waveform displayed ⚫ on the screen will also change, and the total number of lines containing the decoding information in the event table will also be changed.
  • Page 252 Chapter 12 Protocol Decoding RIGOL data, it indicates that the decoded data are not fully displayed. At this time, you can view the details in the "Packets" view. Export: If you select "Packets", you can export time, the corresponding ➢...
  • Page 253 Chapter 12 Protocol Decoding RIGOL related to the value of the horizontal time base) is dispalyed. The detected ACK is 1 MSO8000 User Guide 12-19...
  • Page 254: Spi Decoding (Option)

    Chapter 12 Protocol Decoding RIGOL SPI Decoding (Option) SPI bus is based on the master-slave configuration and usually consists of chip select line (CS), clock line (SCLK), and data line (SDA). Wherein, the data lines include the master input/slave output (MISO) data line and master output/slave input (MOSI) data line.
  • Page 255 Chapter 12 Protocol Decoding RIGOL Press CS and rotate the multifunction knob to select the desired ➢ channel, and then press down the knob to select it. You can also press CS continuously or enable the touch screen to select it. The analog channel (CH1-CH4) and digital channel (D0-D15) can all be selected as the CS channel.
  • Page 256 Chapter 12 Protocol Decoding RIGOL 5. Data setting Press Settings to enter the data setting menu. Endian setting ⚫ Press Endian to select "LSB" or "MSB". By default, it is "MSB". LSB: indicates Least Significant Bit transmission sequence, i.g. the lowest bit of the data is transmitted first.
  • Page 257 Chapter 12 Protocol Decoding RIGOL MOSI). Note: When the source channel of MISO or MOSI is set to "OFF", information of the data line is not displayed in the data table. Open or close the event table: Press Event Table continuously to ➢...
  • Page 258 Chapter 12 Protocol Decoding RIGOL and error information (MISO and/or MOSI) are displayed in the event table. If you select "Details", the detailed data in the specified row will be displayed in the event table. If "…" appears in the row of data, it indicates that the decoded data are not fully displayed.
  • Page 259: Lin Decoding (Option)

    Chapter 12 Protocol Decoding RIGOL LIN Decoding (Option) The oscilloscope samples the LIN signal, and judges each data point to be logic "1" or logic "0" according to the preset threshold level. The LIN decoding is required to specify the LIN signal protocol version.
  • Page 260 Chapter 12 Protocol Decoding RIGOL Set the parity bit ⚫ Press Parity bit continuously to select whether the data contains the parity bit. Set the protocol version ⚫ Press Version and rotate the multifunction knob to select a protocol version that matches the LIN bus signal. Then, press down the knob to select it.
  • Page 261 Chapter 12 Protocol Decoding RIGOL Line Number Time Frame ID Decoded Data CRC Error Event Table LIN Bus Figure 12-9 LIN Decoding Event Table Note: When you adjust the horizontal time base, the waveform displayed ⚫ on the screen will also change, and the total number of lines containing the decoding information in the event table will also be changed.
  • Page 262 Chapter 12 Protocol Decoding RIGOL time, you can view the details in the "Packets" view. Export: If you select "Packets", you can export time and the ➢ corresponding decoded data. Press Export, the save setting menu is displayed. You can export the data table of the packets to the internal or external USB storage device (when detected) in CSV format.
  • Page 263 Chapter 12 Protocol Decoding RIGOL Parity Error Checksum Error ⚫ If the checksum error is detected, CRC is displayed in red, as shown in the figure below. Checksum Error Sync Error ⚫ If the sync error is detected, SYNC is displayed in red, as shown in the figure below.
  • Page 264 Chapter 12 Protocol Decoding RIGOL Sync Error 12-30 MSO8000 User Guide...
  • Page 265: Can Decoding (Option)

    Chapter 12 Protocol Decoding RIGOL CAN Decoding (Option) The oscilloscope samples the CAN signal at the specified sample position. The oscilloscope will also judge each data point (logic "1" or logic "0") according to the preset threshold level. You need to specify the CAN signal type and sample position for CAN decoding.
  • Page 266 Chapter 12 Protocol Decoding RIGOL to the CAN_H bus signal and connect the negative lead to the CAN_L bus signal. Specify the standard signal rate ⚫ Three setting methods are available for baud rate: Press Baud, then set the user-defined baud rate with the displayed ➢...
  • Page 267 Chapter 12 Protocol Decoding RIGOL range is from 10% to 90%. 5. Display-related Setting Press Display to enter the display setting menu. Set the display format ⚫ Press Format, rotate the multifunction knob to select the display format of the bus data. Then, press down the knob to select it. You can also press Format continuously or enable the touch screen to select it.
  • Page 268 Chapter 12 Protocol Decoding RIGOL Line Number Time Frame ID DLC Decoded Data Event Table CAN Bus Figure 12-11 CAN Decoding Event Table Note: When you adjust the horizontal time base, the waveform displayed ⚫ on the screen will also change, and the total number of lines containing the decoding information in the event table will also be changed.
  • Page 269 Chapter 12 Protocol Decoding RIGOL time, you can view the details in the "Packets" view. Export: If you select "Packets", you can export time and the ➢ corresponding decoded data. Press Export, the save setting menu is displayed. You can export the data table of the packets to the internal or external USB storage device (when detected) in CSV format.
  • Page 270: Flexray Decoding (Option)

    Chapter 12 Protocol Decoding RIGOL FlexRay Decoding (Option) FlexRay is a type of differential serial bus configured with three consecutive segments (i.g. header, payload, and trailer). The oscilloscope samples the FlexRay signal at the specified sample position and judges each data point as logic "1" or logic "0"...
  • Page 271 Chapter 12 Protocol Decoding RIGOL desired baud rate. Set the signal type ⚫ Press Signal and rotate the multifunction knob to select a signal type that matches the actual FlexRay bus signal. Press down the knob to select it. You can also press Signal continuously or enable the touch screen to select it.
  • Page 272 Chapter 12 Protocol Decoding RIGOL upper-left side of the bus (when the bus display is enabled). Event table ⚫ Press Event Table to enter the event table setting menu. The event table displays the detailed decoding information in time order in the form of a table, so as to better observe the long decoded data.
  • Page 273 Chapter 12 Protocol Decoding RIGOL Set the view type: Press View, then rotate the multifunction knob ➢ to select the form of the event table to be viewed. Press down the knob to select it. You can also press View continuously or enable the touch screen to select it.
  • Page 274 Chapter 12 Protocol Decoding RIGOL TSS Sync Frame Frame ID PL HCRC CYC Data TCRC 12-40 MSO8000 User Guide...
  • Page 275: I2S Decoding (Option)

    Chapter 12 Protocol Decoding RIGOL I2S Decoding (Option) The oscilloscope samples the I2S signal and judges each data point as logic "1" or logic "0" according to the preset threshold level. I2S decoding is required to specify the serial clock, channel signal, and the data's source channel. You need to set Alignment, WS Low, and other parameters.
  • Page 276 Chapter 12 Protocol Decoding RIGOL keypad to set the threshold of the WS signal. When you modify the WS threshold, a dotted line displaying the current threshold level is displayed on the screen. The dotted line disappears in about 2 s after you stop modifying the threshold.
  • Page 277 Chapter 12 Protocol Decoding RIGOL Press Display to enter the display setting menu. Set the display format ⚫ Press Format, rotate the multifunction knob to select the display format of the bus data. Then, press down the knob to select it. You can also press Format continuously or enable the touch screen to select it.
  • Page 278 Chapter 12 Protocol Decoding RIGOL Line Number Time Left Channel Data Right Channel Data Event Table I2S Bus Figure 12-14 I2S Decoding Event Table Note: When you adjust the horizontal time base, the waveform displayed ⚫ on the screen will also change, and the total number of lines containing the decoding information in the event table will also be changed.
  • Page 279 Chapter 12 Protocol Decoding RIGOL data, it indicates that the decoded data are not fully displayed. At this time, you can view the details in the "Packets" view. Export: If you select "Packets", you can export time and the ➢...
  • Page 280: 1553B Decoding (Option)

    Chapter 12 Protocol Decoding RIGOL 1553B Decoding (Option) The oscilloscope samples the 1553B signal, and judges each data point to be logic "1" or logic "0" according to the preset threshold level. 1553B decoding is required to specify the data channel source and the threshold.
  • Page 281 Chapter 12 Protocol Decoding RIGOL side of the bus (when the bus display is enabled). Event table ⚫ Press Event Table to enter the event table setting menu. The event table displays the detailed decoding information in time order in the form of a table, so as to better observe the long decoded data.
  • Page 282 Chapter 12 Protocol Decoding RIGOL continuously or enable the touch screen to select it. The available formats include "Hex", "Dec", "Bin", and "ASC". Set the view type: Press View, then rotate the multifunction knob ➢ to select the form of the event table to be viewed. Press down the knob to select it.
  • Page 283 Chapter 12 Protocol Decoding RIGOL C/S Word C/S Data Parity Bit Data Word Data Word Data Parity Bit MSO8000 User Guide 12-49...
  • Page 285: Chapter 13 Reference Waveform

    Chapter 13 Reference Waveform RIGOL Chapter 13 Reference Waveform MSO8000 series oscilloscope provides 10 reference waveform positions (Ref1-Ref10). In the actual test process, you can compare the signal waveform with the reference waveform to locate the failure. Contents in this chapter: To Enable Ref Function ◼...
  • Page 286: To Enable Ref Function

    Chapter 13 Reference Waveform RIGOL To Enable Ref Function Press Ref on the front panel to enable the reference waveform function, or enable the touch screen and then tap the function navigation icon at the lower-left corner of the screen to open the function navigation. Then, tap the "Ref" icon to enable the reference waveform function.
  • Page 287: To Save To Internal Memory

    Chapter 13 Reference Waveform RIGOL Wave Vertical SCALE and Wave Vertical POSITION to adjust the vertical scale and vertical offset. To Save to Internal Memory Press SaveToRef to save the displayed waveform for the specified source to the internal memory as the reference waveform.
  • Page 288: To Reset The Reference Waveform

    Chapter 13 Reference Waveform RIGOL To Reset the Reference Waveform Press More → Reset, and the reference waveform is restored to the position where the source channel waveform is located when the SaveToRef operation was executed. Color Setting MSO8000 series oscilloscope provides five colors (gray, green, blue, red, and orange) to mark the reference waveforms of different channels in order to distinguish the reference waveforms of different reference waveform channels.
  • Page 289: To Export To Internal Or External Memory

    Chapter 13 Reference Waveform RIGOL To Export to Internal or External Memory You can also save the current reference waveform to the internal memory or external USB storage device. The file format of the reference waveform is ".ref", ".bin", or ".csv".
  • Page 291: Chapter 14 Pass/Fail Test

    Chapter 14 Pass/Fail Test RIGOL Chapter 14 Pass/Fail Test During the product design and manufacturing process, you usually need to monitor the variations of the signal or judge whether the product is up to standard. The pass/fail test function of MSO8000 series oscilloscope can accomplish this task perfectly.
  • Page 292: To Enable Or Disable The Pass/Fail Test Function

    Chapter 14 Pass/Fail Test RIGOL To Enable or Disable the Pass/Fail Test Function Press Utility → PassFail to open the pass/fail test setting menu, or enable the touch screen and then tap the function navigation icon at the lower-left corner of the screen to open the function navigation.
  • Page 293: To Create A Mask

    Chapter 14 Pass/Fail Test RIGOL To Create a Mask Press Mask → Create to self-define the mask of the pass/fail test. Press Range and rotate the multifunction knob to select the mask range. Press down the knob to select it. You can also press Range continuously or enable the touch screen to select it.
  • Page 294: To Set The Output Form Of The Test Results

    Chapter 14 Pass/Fail Test RIGOL them to the current pass/fail test function. To Set the Output Form of the Test Results Press Option to enter the "Option" menu. You can set what the oscilloscope will execute when test results are detected according to your needs.
  • Page 295: To Enable Or Disable The Display Of The Statistics Of The Test Results

    Chapter 14 Pass/Fail Test RIGOL To Enable or Disable the Display of the Statistics of the Test Results Press Information continuously to enable or disable the display of the statistics of the test results. You can also enable the touch screen to tap the icon at the right-upper corner of the statistics window to close it.
  • Page 297: Chapter 15 Waveform Recording & Playing

    Chapter 15 Waveform Recording & Playing RIGOL Chapter 15 Waveform Recording & Playing Waveform recording & playing function can play the recorded waveforms of the analog input channels (CH1-CH4) and the digital channels (D0-D15), enabling you to analyze the waveforms better.
  • Page 298: Common Settings

    Chapter 15 Waveform Recording & Playing RIGOL Common Settings Press Utility → Record to enter the waveform recording function setting menu. You can also enable the touch screen and then tap the function navigation icon the lower-left corner of the screen to open the function navigation. Then, tap the "Record"...
  • Page 299: Record Options

    Chapter 15 Waveform Recording & Playing RIGOL 4. Set Transition Press Jump To continuously to select to jump to the recorded end frame Last) or the recorded start frame ( First). For settings of the start and end frames, refer to descriptions in "Play Options".
  • Page 300: Play Options

    Chapter 15 Waveform Recording & Playing RIGOL Memory Depth = Sample Rate x Horizontal Time Base x Number of Grids in the Horizontal Direction. Therefore, the maximum number of frames of waveform recording is also related to the "Sample Rate" and "Horizontal Time Base". This oscilloscope provides ceaseless recording for a maximum of 450,000 frames of real-time waveforms.
  • Page 301 Chapter 15 Waveform Recording & Playing RIGOL 5. End Frame Press End Frame, then rotate the multifunction knob or use the pop-up numeric keypad to set the end frame of playback. The default is the number of frames of the recorded waveforms.
  • Page 303: Chapter 16 Search And Navigation Function

    Chapter 16 Search and Navigation Function RIGOL Chapter 16 Search and Navigation Function The search function allows you to search for relevant events based on the search condition that you set. The navigation function includes the recording & playing navigation, time navigation, and event navigation.
  • Page 304: Search Function

    Chapter 16 Search and Navigation Function RIGOL Search Function The search function enables you to search the channel edge, Pulse trigger, Runt trigger, Slope trigger, RS232 trigger, I2C trigger, or SPI trigger. Press Search on the front panel to enable the search setting menu. You can also enable the touch screen...
  • Page 305 Chapter 16 Search and Navigation Function RIGOL knob or use the pop-up numeric keypad to set the thresholds of SCL and SDA. SPI: Press Type to select "SPI" as the search type. For SPI Setup menu ⚫ settings, refer to descriptions in "SPI Trigger (Option)". Press Threshold to enter the threshold setting menu.
  • Page 306: Navigation Function

    Chapter 16 Search and Navigation Function RIGOL navigation combination keys to make navigations at a step of 1. For details, refer to descriptions in "Navigation Function". 6. Save You can save the event mark data to the internal memory or external USB storage device in "*.csv"...
  • Page 307 Chapter 16 Search and Navigation Function RIGOL Event Navigation ⚫ When you enable the navigation function and complete the event search, you can use the navigation combination keys to quickly navigate the specific event in the event mark table. Press to navigate to the previous event (the serial number in the mark table decreases);...
  • Page 309: Chapter 17 Display Control

    Chapter 17 Display Control RIGOL Chapter 17 Display Control In the display control setting menu, you can set the parameters such as the waveform display type, persistence time, intensity, grid type, grid brightness, and etc. Contents in this chapter: To Select the Display Type ◼...
  • Page 310: To Select The Display Type

    Chapter 17 Display Control RIGOL To Select the Display Type Press Display to enter the display control setting menu. Press Type continuously to set the waveform display mode to "Vector" or "Dots". Vector display: indicates that the sample points are connected by lines and ⚫...
  • Page 311: To Set The Waveform Intensity

    Chapter 17 Display Control RIGOL To Set the Waveform Intensity Press Intensity, then rotate the multifunction knob at a small step value or use the pop-up numeric keypad to set the waveform intensity. Note that in the non-menu operation mode, you can also rotate the multifunction knob to adjust the waveform brightness of the channel.
  • Page 312: Scale

    Chapter 17 Display Control RIGOL use the pop-up numeric keypad to set the grid intensity. The default is 20%, and the range available is from 1% to 100%. Scale Press Show Scale continuously to enable or disable the scale display on the screen.
  • Page 313: Chapter 18 Function/Arbitrary Waveform Generator (Option)

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL Chapter 18 Function/Arbitrary Waveform Generator (Option) MSO8000 series has an optional built-in dual-channel, 25 MHz signal source, which integrates the signal source and the oscilloscope into one, providing great convenience for engineers who need to use the signal source and oscilloscope at the same time.
  • Page 314: To Output Basic Waveforms

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL To Output Basic Waveforms Press GI on the front panel or enable the touch screen to tap the GI label at the bottom of the screen to enable/disable the output of the [GI] connector on the front panel, and enter the Function/Arbitrary Waveform Generator setting interface.
  • Page 315: To Output Square

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL continuously to select it. Also, you can enable the touch screen to select it. Rotate the multifunction knob or use the pop-up numeric keypad to set the offset or low level value of the current signal. When Impedance is set to "HighZ", the available range is from (-2.5 V+currently set amplitude value/2) to (2.5 V-...
  • Page 316: To Output Ramp

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL To Output Ramp Press Wave, rotate the multifunction knob to select "Ramp". Press down the knob to select it. You can also press Wave continuously or enable the touch screen to select it. At this time, you can set the parameters for the ramp waveform. For setting methods, refer to descriptions in "To Output Sine".
  • Page 317: To Output Dc

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL Press Settings → Duty Cycle, then rotate the multifunction knob or use the pop-up numeric keypad to set the duty cycle of the current pulse signal. Its settable range is from 10% to 90%.
  • Page 318: Sinc

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL offset or low level value of the current signal. 3. Set the impedance Press Impedance continuously to set the output impedance of the Function/Arbitrary Waveform Generator. It can be set to "HighZ" or "50Ω".
  • Page 319: Exp.fall

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL Exp.Fall Press Wave, rotate the multifunction knob to select "Exp.Fall". Press down the knob to select it. You can also press Wave continuously or enable the touch screen to select it. At this time, you can set the parameters for the Exp.Fall waveform (as shown in the figure below).
  • Page 320: Lorentz

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL knob to select it. You can also press Wave continuously or enable the touch screen to select it. At this time, you can set the parameters for the Gauss waveform (as shown in the figure below). For setting methods, refer to descriptions in "To Output Sine".
  • Page 321: To Output The Arbitrary Waveform

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL in "To Output Sine". Figure 18-9 Haversine To Output the Arbitrary Waveform MSO8000 series oscilloscope allows you to define the arbitrary waveform and save it to the internal or external memory. The user-defined waveform can contain 2 to 16,384 data points (i.g.
  • Page 322: To Create The Waveform

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL range. Load ⚫ Press Load to load the set channel signal. 2. Load the Waveform Press Load Stored to enter the waveform loading interface. In the disk management interface, select the waveform from the internal memory or the external memory.
  • Page 323 Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL Set the initial number of points ⚫ The initial number of points indicates the number of editable points. By default, the number of editable points of arbitrary waveforms created can be preset to 2.
  • Page 324: To Edit Waveforms

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL external memory in "*.arb" format (you can overwrite the original file or save the currently edited waveform again). To Edit Waveforms You can edit the waveforms in the current volatile memory. Press Edit to enter the waveform editing menu.
  • Page 325: Modulation

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL Current Point ➢ Press PointX, then rotate the multifunction knob or use the pop-up numeric keypad to select the points to be edited. The available range is from 1 to the number of initial points.
  • Page 326 Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL type, and then press down the knob to select it. You can also press Type continuously or enable the touch screen to select it. The available modulation types include AM, FM, and FSK.
  • Page 327 Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL is half of the carrier amplitude. When it is set to 100%, the output amplitude equals the carrier amplitude. When it is set to a value greater than 100%, envelop distortion will occur, which must be avoided in actual circuit; at this point, the output of the instrument will not exceed 2.5 Vpp (the load is 50 Ω).
  • Page 328: Fsk

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL of the current carrier waveform. FSK (Frequency-shift Keying) is a modulation method that uses the digital signal to control the carrier frequency changes. It is one of the modulation methods that have been early used in information transmission. Its advantages lie in good performance in anti-noise and anti-attenuation, and it has been widely used in low and medium rate of data transmission.
  • Page 329 Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL Log: the frequency of the signal changes in log form. ➢ Step: the frequency of the signal changes with ladder-like step. ➢ 2. Sweep Settings Press Settings to enter the sweep setting menu.
  • Page 330: Burst

    Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL Set the step ⚫ When the sweep type is set to "Step", this menu is enabled and can be settable. Press Step number, then rotate the multifunction knob use the pop-up numeric keypad to set the step value. By default, it is 2. Its available range is from 2 to 1000.
  • Page 331 Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL 1. Set the Burst Type Press Burst Type, and rotate the multifunction knob to select the desired burst type, and then press down the knob to select it. You can also press Burst Type continuously or enable the touch screen to select it.
  • Page 332 Chapter 18 Function/Arbitrary Waveform Generator (Option) RIGOL It is defined as the time from the beginning of one burst to that of another burst. Press Burst Period, then rotate the multifunction knob or use the pop-up numeric keypad to set the desired burst period. Its available range is from 2 ms to 500 s.
  • Page 333: Chapter 19 Store And Load

    Chapter 19 Store and Load RIGOL Chapter 19 Store and Load You can save the current settings, waveforms, screen image, and parameters of the oscilloscope in internal memory or external storage device (such as USB storage device) in various formats and load the stored settings or waveforms when needed.
  • Page 334: Storage System

    Chapter 19 Store and Load RIGOL Storage System Press Storage on the front panel to enter the store and load setting interface. It provides 3 USB HOST interfaces on the front panel and one USB HOST interface on the rear panel, which can all be connected to the USB storage device for external storage.
  • Page 335 Chapter 19 Store and Load RIGOL 2. Wave Press Save Wave to enter the waveform storage setting interface. Save the main setting information (e.g. "On/Off" state of the channel, vertical scale, and horizontal time base) and waveform data of all enabled channels (analog channel and digital channel) to the external memory.
  • Page 336: Binary Data Format (.Bin)

    1. File Header There is only one file header in a binary file. The file header contains the following information. Cookie Two-byte characters, RG, indicating that the file is the RIGOL binary data file format. Version Two bytes, indicating the file version.
  • Page 337 Chapter 19 Store and Load RIGOL 2 = Peak Detection 3 = Average 4 = Not Used 5 = Not Used 6 = Logic Number of A 4-byte integer, indicating the number of waveform buffers Waveform required to read the data.
  • Page 338: Load Type

    Chapter 19 Store and Load RIGOL above under X Units. Date A 16-byte character array, not used Time A 16-byte character array, not used Frame A 24-byte character array, indicating the model number and serial number of the oscilloscope in the format: MODEL#:SERIAL#.
  • Page 339: Internal Storage And Load

    Chapter 19 Store and Load RIGOL Press Load Setup to load the setups from the internal memory or the external USB disk. The default file type is "*.stp", and no other options are available. Load the file from the memory, then press Load to load the selected file.
  • Page 340 Chapter 19 Store and Load RIGOL management interface is displayed automatically. By default, "Local Disk(C)" is selected. 2) Rotate the multifunction knob or enable the touch screen to open the internal memory storage interface. 3) Rotate the multifunction knob to select the desired storage location, and then press down the knob to open it.
  • Page 341: External Storage And Load

    Chapter 19 Store and Load RIGOL External Storage and Load Before external storage and recall, ensure that the USB disk is properly connected. The external memory supports image saving, waveform saving, setup saving, waveform loading, and setup loading. 1. Save the Specified Type of File in the External USB Storage Device...
  • Page 342: To Select A File Type

    Chapter 19 Store and Load RIGOL Figure 19-1 Disk Management Interface Execute the following operations through the disk management menu: To Select a File Type ⚫ To Create a Folder ⚫ To Delete a File or Folder ⚫ To Copy and Paste a File or Folder ⚫...
  • Page 343 Chapter 19 Store and Load RIGOL Figure 19-2 To Create a Folder This oscilloscope supports Chinese/English input method. The length of the folder name is limited to 22 bytes (one Chinese character occupies two bytes). The following part introduces how to input a folder name using Chinese/English input method.
  • Page 344 Chapter 19 Store and Load RIGOL 1. Select English Input Method If the current input method is "En/中/繁", please go to Step 2. If the current input method is "中/繁/En" or "繁/En/中", rotate the multifunction knob select the input switch key and press the knob continuously to switch to "En/中/繁".
  • Page 345 Chapter 19 Store and Load RIGOL Simplified Chinese (or Traditional Chinese) Input Method For example, create a folder with the name "文件名". Note that the input method of traditional Chinese is the same as that of the simplified Chinese. Pinyin Input Area Chinese Character Selecting Area 1.
  • Page 346: To Delete A File Or Folder

    Chapter 19 Store and Load RIGOL Refer to Step 3 and use the same method to input the remaining Chinese characters "件名" in sequence. 5. Modify or Delete the Unwanted Characters that Have Been Entered. When inputting the filename, you can modify or delete the filename characters input in the "Name Input Area"...
  • Page 347: To Copy And Paste A File Or Folder

    Chapter 19 Store and Load RIGOL "Yes" to delete the file or the folder. When you delete a directory where a file or a folder exists, the directory ⚫ cannot be deleted. The touch screen function only allows you to select the directory and open it.
  • Page 348: Factory Settings

    Chapter 19 Store and Load RIGOL memory "Local Disk(C)". Press SecurityClear, and then a prompt message for confirming security clear is displayed. Press OK to clear all the files stored in the internal memory. Factory Settings Press Default on the front panel, then a prompt message "Restore default?" is displayed.
  • Page 349 Chapter 19 Store and Load RIGOL Channel Coupling BW Limit Attenuation Input Impedance 1 MΩ Invert Fine Channel Unit Display Label Built-in Label Ch-Ch Skew Offset Cal Acquire Acquisition Mode Normal Memory Depth Auto Anti-Aliasing Trigger Trigger Type Edge Trigger...
  • Page 350 Chapter 19 Store and Load RIGOL Frequency 1 kHz Amplitude 500 mV Offset Start Phase 0° Setting Type Impedance HighZ Cursor Mode Manual Select Source 100 mV -100 mV Vertical Unit Source Hori. Unit Region Main Track AX Source BX Source -1 μs...
  • Page 351 Chapter 19 Store and Load RIGOL Save Setup File Type *.stp Window Show Load Wave File Type *.wfm Load Setup File Type *.stp Disk File Type Utility Sound PassFail Enable Source Operate Stop Range Screen X Mask 0.24 div Y Mask 0.48 div...
  • Page 352 Chapter 19 Store and Load RIGOL Jump To First Save First System Power On Default Power status Switch On Aux Out TrigOut Key Locker Unlocked HDMI Output HDMI Resolution 1280x720 SelfCal Window Open Auto Config Peak to Peak Overlay Coupling...
  • Page 353 Chapter 19 Store and Load RIGOL SourceA SourceB Offset Scale 500 mV A× B Operation SourceA SourceB Offset Scale 500 mU A÷B Operation SourceA SourceB Offset Scale 500 mU Operation Source Offset 0 dBV Scale 2 dBV Unit dBm/dBV Start-End...
  • Page 354 Chapter 19 Store and Load RIGOL Operation SourceA SourceB Wave Size Medium Thre.CH1 Thre.CH2 Thre.CH3 Thre.CH4 Sensitivity 0.3 Div Operation SourceA SourceB Wave Size Medium Thre.CH1 Thre.CH2 Thre.CH3 Thre.CH4 Sensitivity 0.3 Div Operation SourceA Wave Size Medium Thre.CH1 Thre.CH2 Thre.CH3 Thre.CH4...
  • Page 355 Chapter 19 Store and Load RIGOL Source Offset Scale 500 mU Invert Operation Source Offset Scale 500 mU Invert Operation Source Offset Scale 500 mU Invert Operation Source Offset Scale 500 mU Invert Operation Source Offset Scale 500 mV Invert...
  • Page 356 Chapter 19 Store and Load RIGOL Source Offset Scale 500 mV ωc1 50 MHz ωc2 100 MHz Invert Band Stop Operation Source Offset Scale 500 mV ωc1 50 MHz ωc2 100 MHz Invert AX+B Operation Source Offset Scale 500 mV...
  • Page 357 Chapter 19 Store and Load RIGOL Protocol Decoding (Decode) Bus Type Parallel Bus Status Format Label Event Table Noise Reject Clock RS232 Bus Status Baud 9.6 kbps Threshold Polarity Endian Data 8 bits Stop Bit 1 bit Parity None Package...
  • Page 358 Chapter 19 Store and Load RIGOL Source Threshold Baud 19.2 kbps Parity Without Version Both Bus Status Source Threshold Signal CAN_L Baud 1 Mbps Sample Position FlexRay Bus Status Source Threshold Channel Selection Baud 10 Mbps Signal Sample Position Bus Status...
  • Page 359 Chapter 19 Store and Load RIGOL VScale VOffset Color Orange Label Display Library Default Label REF1 Note : This function is only available for the model installed with the MSO8000-AWG option. MSO8000 User Guide 19-27...
  • Page 361: Chapter 20 System Utility Function Setting

    Chapter 20 System Utility Function Setting RIGOL Chapter 20 System Utility Function Setting In the system function setting menu, you can set the parameters for the remote interface and the system-related parameters. Contents in this chapter: Remote Interface Configuration ◼...
  • Page 362: Remote Interface Configuration

    Remote Interface Configuration MSO8000 series can communicate with the PC via the USB, LAN, or GPIB (requiring to work with RIGOL's USB-GPIB interface converter) interface. Before using the remote interfaces, configure the corresponding interface according to the introductions in the following section.
  • Page 363 Chapter 20 System Utility Function Setting RIGOL Connected! ⚫ Invalid IP! ⚫ IP lost! ⚫ Please wait... ⚫ IP Configuration Type The configuration type of the IP address can be DHCP, Auto IP, or Static IP. For different IP configuration types, the configurations for IP address and other network parameters are different.
  • Page 364 Chapter 20 System Utility Function Setting RIGOL input the desired IP address. This setting will be saved to the non-volatile memory; if "Power On" is set to "Last", then DHCP and Auto IP are disabled at the next power-on. The oscilloscope will load the preset IP address automatically.
  • Page 365: To Set Mdns

    Chapter 20 System Utility Function Setting RIGOL When the three IP configuration types are all turned on, the priority of the ⚫ parameter configuration from high to low is "DHCP", "Auto IP", and "Static IP". The three IP configuration types cannot be all turned off at the same time.
  • Page 366: To Set The Gpib Address

    Chapter 20 System Utility Function Setting RIGOL To Set the GPIB Address Before using the GPIB bus to control the oscilloscope, use the USB-GPIB interface converter (an option that requires to be ordered separately) to extend the GPIB interface, and then use the GPIB cable to connect the instrument to the PC. Press Utility →...
  • Page 367: System-Related

    Chapter 20 System Utility Function Setting RIGOL System-related Beeper When the beeper is enabled, you can hear the sound of the beeper when you perform the following operations: Press a key or a menu key on the front panel ⚫...
  • Page 368: Power Status

    Chapter 20 System Utility Function Setting RIGOL on again after power-off. Press Utility → System, and then press Power On continuously to select "Last" or "Default". By default, it is "Default". Last: returns to the setting of the system at last power-off.
  • Page 369: Ref Clock

    Monday you power on the instrument, the oscilloscope will check whether the latest version of the upgrade software is available on the RIGOL official website. If yes, a red spot appears at the upper-left corner of the Online upgrade menu, prompting you that the latest upgrade version is available and you can perform online upgrade.
  • Page 370: Self-Calibration

    Chapter 20 System Utility Function Setting RIGOL to insert the USB disk that contains the software installation package); tap Cancel to cancel local upgrading operation. Self-calibration The self-calibration program can quickly make the oscilloscope to work in an optimal state to get the precise measurement results. You can perform self-calibration at any time, especially when the changes of the ambient temperature reach or above 5℃.
  • Page 371: Auto Config

    Chapter 20 System Utility Function Setting RIGOL Auto Config You can set the parameters for the AUTO menu. Press Utility → More → Auto Config, and set the parameters for the AUTO key. For detailed settings, refer to descriptions in "Quick Measurement after AUTO".
  • Page 372: Email

    Chapter 20 System Utility Function Setting RIGOL Press Print test page to carry out the print test. The print progress is also displayed on the screen. 2. Set the Printer Parameters Print ⚫ After the printer parameters are set, press Print to execute the print operation.
  • Page 373: Key Locker

    Chapter 20 System Utility Function Setting RIGOL Set SMTP ⚫ Press SMTP to set the mail transmission protocol. For detailed settings of mail transmission protocol, refer to descriptions in "To Create a Folder". Set the port ⚫ Press Port to set the protocol port in the pop-up numeric keypad. Its range is from 1 65535.
  • Page 374: Quick Operation

    Chapter 20 System Utility Function Setting RIGOL Locked: indicates that all keys of the oscilloscope (except the Key Locker key) ⚫ are locked. Unlocked: unlocks the front-panel keys. ⚫ Quick Operation You can set the shortcut functions for the Quick key on the front panel.
  • Page 375 Chapter 20 System Utility Function Setting RIGOL Press Operation and then rotate the multifunction knob to select "Save Wave". Press down the knob to select it. You can also press Operation continuously to select among the options, or enable the touch screen to tap the desired operation type.
  • Page 376 Chapter 20 System Utility Function Setting RIGOL desired measurement channel. You can also press All Measure continuously to select or enable the touch screen to tap the desired channel. The available measurement channels include OFF and CH1-CH4. 5. Statistics Reset When the Quick shortcut key is set to "Stat Reset", press Quick on the front...
  • Page 377 The file types include "*.png", "*.bmp", "*.jpg", and "*.tif". If you select "Text" under Screen Saver, press Text, and then the ⚫ filename input interface is displayed, e.g. inputting "RIGOL Scope". For details, refer to descriptions in "To Create a Folder". 2. Set the Wait Time...
  • Page 378: Self-Check

    Chapter 20 System Utility Function Setting RIGOL 3. Preview Press Preview to view the set screen saver. 4. Default Press Default to restore the screen saver to the default settings. Self-check The oscilloscope supports a variety of self-check functions, including key test, screen test, and touch test.
  • Page 379: System Time

    The press SINGLE to switch to the next touch screen test interface, as shown in the figure below. At this time, you can pinch or stretch the RIGOL logo on the screen to zoom it out or in to check whether the gestures work normally. Press RUN/STOP for three consecutive times to exit the touch screen test interface.
  • Page 380: Default Option

    Chapter 20 System Utility Function Setting RIGOL 2. Apply the Currently Set System Time Press Apply to validate the current settings. The time at the lower-right side of the screen will be updated. Default Option Used for configuring the Default key.
  • Page 381: Chapter 21 Remote Control

    Remote Control via GPIB ◼ This chapter will illustrate how to use RIGOL's Ultra Sigma software to remotely control the oscilloscope via various interfaces. Note: When communicating with the PC via GPIB, the oscilloscope does not support large data transmission operation such as screen shot and waveform reading.
  • Page 382: Remote Control Via Usb

    Chapter 21 Remote Control RIGOL Remote Control via USB 1. Connect the device Use a USB cable to connect the oscilloscope (USB DEVICE interface) with the PC (USB HOST interface). 2. Install the USB driver This oscilloscope is a USB-TMC device. After you connect the oscilloscope to the PC properly and power it on for the first time (the oscilloscope will be automatically configured with the USB interface).
  • Page 383 Chapter 21 Remote Control RIGOL MSO8000 User Guide 21-3...
  • Page 384 PC via the USB interface. You can also click to search for the resource. 4. View the device resource The searched instrument resource will be under the directory of "RIGOL Online Resource", with the model of the instrument and the USB interface information being displayed as follows.
  • Page 385 Chapter 21 Remote Control RIGOL 5. Control the instrument remotely Right-click the resource name "MSO8104 (USB0::0x1AB1::0x0516::DS8F201200055::INSTR)" and select "SCPI Panel Control" to turn on the remote command control panel, through which you can send commands and read data. MSO8000 User Guide...
  • Page 386: Remote Control Via Lan

    4. View the device resource The resources found will appear under the "RIGOL Online Resource" directory. For example, MSO8104 (TCPIP::172.16.3.14::INSTR). 5. Control the instrument remotely Right-click the resource name "MSO8104 (TCPIP::172.16.3.14::INSTR)"...
  • Page 387: Remote Control Via Gpib

    Chapter 21 Remote Control RIGOL select "LXI-Web"). Various important information about the instrument (including the model, manufacturer, serial number, description, MAC address, and IP address) will be displayed on the webpage. You can also directly input the IP address of the oscilloscope in the address bar of the PC browser to load the LXI webpage.
  • Page 388 5. View the device resource The searched instrument resource will be under the directory of "RIGOL Online Resource", as shown in the following figure. 6. Control the instrument remotely Right-click the resource name "MSO8104 (GPIB0::18::INSTR)", and select "SCPI...
  • Page 389: Chapter 22 Troubleshooting

    The commonly encountered failures of the oscilloscope and their solutions are listed below. If the following problems occur, locate and resolve the problems according to the following steps. If the problems still persist, contact RIGOL and provide your instrument information to us.
  • Page 390 (3) Check whether there is a strong magnetic field around the instrument. If the instrument is close to the strong magnetic field (e.g. a magnet), please move the instrument away from the magnet field. (4) If the problem still persists, please contact RIGOL. 22-2 MSO8000 User Guide...
  • Page 391: Chapter 23 Appendix

    Model MSO8204 (2 GHz, 10 GSa/s, 500 Mpts, 4+16 CH MSO) MSO8204 MSO8104 (1 GHz, 10 GSa/s, 500 Mpts, 4+16 CH MSO) MSO8104 MSO8064 (600 MHz, 10 GSa/s, 500 Mpts, 4+16 CH MSO) MSO8064 Standard Accessories CB-USBA-USBB-FF-1 USB cable 4 passive high-impedance probes (500 MHz) RP3500A 2 passive low-impedance probes (1.5 GHz, only for the...
  • Page 392 Dual-channel 25 MHz arbitrary waveform generator MSO8000-AWG Built-in power analysis (required to purchase the RPA246 MSO8000-PWR phase deviation correction jig) Real-time eye diagram and jitter analysis MSO8000-JITTER Note: For all the accessories and options, please contact the local office of RIGOL. 23-2 MSO8000 User Guide...
  • Page 393: Appendix B: Warranty

    If a product proves defective within the warranty period, RIGOL guarantees free replacement or repair for the defective product. To get repair service, please contact with your nearest RIGOL sales or service office. There is no other warranty, expressed or implied, except such as is expressly set forth herein or other applicable warranty card.
  • Page 395: Index

    Index RIGOL Index +Duty ........6-33 Hanning ........6-9 +Rate ........6-37 Haversine ......... 18-9 +Width ......... 6-32 Horizontal Expansion ....4-11 Acquisition Mode ......4-4 Horizontal Position ...... 3-3 AM ........18-14 I2C Decoding ......12-15 Anti-Aliasing ......4-11 I2C Trigger .......
  • Page 396 Index RIGOL Pulse Width Setting ..... 5-11 Trigger Holdoff ......5-6 Qualifier ....... 5-24 Trigger LEVEL/Threshold Level ..5-3 Ramp ........18-4 Trigger Mode ......5-4 -Rate ........6-37 Trigger Position ....5-26 Rectangular ....... 6-9 tVmax ........6-33 Ref Clock ......... 20-9 tVmin ........

This manual is also suitable for:

Mso8000 seriesMso8104Mso8204

Table of Contents