Load Type; Internal Storage And Load - Rigol MSO8000A Series User Manual

Digital oscilloscope
Hide thumbs Also See for MSO8000A Series:
Table of Contents

Advertisement

Header Size
Buffer Type
Bytes Per Point
Buffer Size
22.3

Load Type

This oscilloscope provides two load types: waveform load and setup load. The
descriptions of each load type are as follows.
1. Waveform Load
Press Load Wave to load the waveforms from the internal memory or the external
USB disk. The default file type is "*.wfm" and no other options are available. Select
the file to be loaded from the memory, then press Load to load the selected file.
2. Load Setup
Press Load Setup to load the setups from the internal memory or the external USB
disk. The default file type is "*.stp", and no other options are available. Select the file
to be loaded from the memory, then press Load to load the selected file.
22.4

Internal Storage and Load

The internal memory supports the storage and load of the image file, waveform file,
setup file, reference waveform file, and pass/fail test mask file.
1. Save the Specified Oscilloscope Setting to the Internal Memory (Take Save
Setup as an Example)
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
A 4-byte integer, indicating the number of bytes in the
waveform data header.
A 2-byte integer, indicating the type of the waveform data
stored in the file.
0 = Unknown
1 = Normal 32-bit float data
2 = Maximum float data
3 = Minimum float data
4 = Not Used
5 = Not Used
6 = Digital unsigned 8-bit character data (for digital
channels)
A 2-byte short integer, indicating the number of bytes per
data point.
A 4-byte integer, indicating the size of the buffer required
to hold the data points.
Store and Load
329